Suppr超能文献

机器人辅助定制型全膝关节置换术:手术技术与初步结果

Robotic-assisted differential total knee arthroplasty with patient-specific implants: surgical techniques and preliminary results.

作者信息

Zheng Hanlong, Chen Mingxue, Yang Dejin, Shao Hongyi, Zhou Yixin

机构信息

Department of Orthopaedic Surgery, Beijing Jishuitan Hospital, Capital Medical University, No. 31, Xinjiekou East Street, Xicheng District, Beijing, 100035, China.

出版信息

Arthroplasty. 2024 Jun 10;6(1):34. doi: 10.1186/s42836-024-00255-1.

Abstract

OBJECTIVE

In total knee arthroplasty (TKA), achieving soft-tissue balance while retaining acceptable lower limb alignment is sometimes difficult and may lead to patient dissatisfaction. Theoretically, patient-specific implants can bring great benefits, while the lack of precise surgical tools may hinder the improvement of outcomes. The objective of this study was to illustrate surgical techniques and evaluate kinematics and early clinical outcomes of robotic-assisted TKA using patient-specific implants.

METHODS

Based on preoperative CT scan, femoral and tibial components were 3D printed. Medial and lateral tibial liners were separate with different thicknesses, posterior slopes and conformity. TiRobot Recon Robot was used for surgery, and was armed with smart tools that quantify gap, force and femoral-tibial track. We collected data on demographics, intraoperative gap balance and femoral-tibial motion. In the follow-up, we evaluated the range of motion, Visual Analogue Scale (VAS), forgotten joint score (FJS), Knee injury and Osteoarthritis Outcome Score, Joint Replacement (KOOS, JR) score. Radiological data were also harvested.

RESULTS

Fifteen patients (17 knees) were enrolled with a mean age of 64.6 ± 6.4 (53-76) years. In 5 knees, we used symmetric tibial liners, the rest were asymmetric. After surgery, the average alignment was 1.6 ± 2.0 (-3-5) degrees varus. The average follow-up lasted 6.7 ± 4.2 (1-14) months. The mean visual analogue scale was 0.8 ± 0.7 (0-2), FJS was 62.4 ± 25.3 (0-87), KOOS was 86.5 ± 9.4 (57-97). 11 patients were "very satisfied", 3 were "satisfied" with the result, and one patient was neutral due to restricted extension and unsatisfactory rehabilitation at five months' follow-up.

CONCLUSIONS

With patient-specific implants and robotics, TKA could be performed by a mathematical way, which was dubbed a "differential" TKA. Intraoperative kinematics was excellent in terms of gap-force balancing and femoral-tibial relative motion. Preliminary clinical outcomes were overall satisfactory.

摘要

目的

在全膝关节置换术(TKA)中,在保持可接受的下肢对线的同时实现软组织平衡有时很困难,可能导致患者不满意。理论上,定制植入物可以带来巨大益处,但缺乏精确的手术工具可能会阻碍手术效果的改善。本研究的目的是阐述使用定制植入物的机器人辅助TKA的手术技术,并评估其运动学和早期临床结果。

方法

基于术前CT扫描,对股骨和胫骨部件进行3D打印。胫骨内外侧衬垫分开,具有不同的厚度、后倾角度和贴合度。使用TiRobot Recon机器人进行手术,该机器人配备了能够量化间隙、力和股骨-胫骨轨迹的智能工具。我们收集了人口统计学数据、术中间隙平衡和股骨-胫骨运动数据。在随访中,我们评估了活动范围、视觉模拟评分(VAS)、遗忘关节评分(FJS)、膝关节损伤和骨关节炎疗效评分、关节置换(KOOS,JR)评分。还收集了放射学数据。

结果

共纳入15例患者(17个膝关节),平均年龄64.6±6.4(53 - 76)岁。5个膝关节使用了对称胫骨衬垫,其余为不对称衬垫。术后,平均对线为内翻1.6±2.0(-3 - 5)度。平均随访时间为6.7±4.2(1 - 14)个月。平均视觉模拟评分为0.8±0.7(0 - 2),FJS为62.4±25.3(0 - 87),KOOS为86.5±9.4(57 - 97)。11例患者“非常满意”,3例对结果“满意”,1例患者在5个月随访时因伸直受限和康复效果不满意而持中立态度。

结论

使用定制植入物和机器人技术,TKA可以通过数学方式进行,这被称为“差异”TKA。术中在间隙-力平衡和股骨-胫骨相对运动方面的运动学表现出色。初步临床结果总体令人满意。