Anal Chem. 2024 Jun 25;96(25):10237-10245. doi: 10.1021/acs.analchem.4c00837. Epub 2024 Jun 13.
Dot-blot immunoassays are widely used for the user-friendly detection of clinical biomarkers. However, the majority of dot-blot assays have only limited sensitivity and are only used for qualitative or semiquantitative analysis. To overcome this limitation, we have employed labels based on photon-upconversion nanoparticles (UCNPs) that exhibit anti-Stokes luminescence and can be detected without optical background interference. First, the dot-blot immunoassay on a nitrocellulose membrane was optimized for the quantitative analysis of human serum albumin (HSA), resulting in a limit of detection (LOD) of 0.19 ng/mL and a signal-to-background ratio (/) of 722. Commercial quantum dots were used as a reference label, reaching the LOD of 4.32 ng/mL and the / of 3, clearly indicating the advantages of UCNPs. In addition, the potential of UCNP-based dot-blot for real sample analysis was confirmed by analyzing spiked urine samples, reaching the LOD of 0.24 ng/mL and recovery rates from 79 to 123%. Furthermore, we demonstrated the versatility and robustness of the assay by adapting it to the detection of two other clinically relevant biomarkers, prostate-specific antigen (PSA) and cardiac troponin (cTn), reaching the LODs in spiked serum of 9.4 pg/mL and 0.62 ng/mL for PSA and cTn, respectively. Finally, clinical samples of patients examined for prostate cancer were analyzed, achieving a strong correlation with the reference electrochemiluminescence immunoassay (recovery rates from 89 to 117%). The achieved results demonstrate that UCNPs are highly sensitive labels that enable the development of dot-blot immunoassays for quantitative analysis of low-abundance biomarkers.
点印迹免疫分析因其使用方便而被广泛用于临床生物标志物的检测。然而,大多数点印迹分析方法的灵敏度有限,只能用于定性或半定量分析。为了克服这一限制,我们采用了基于上转换纳米粒子(UCNPs)的标记物,其具有反斯托克斯发光特性,无需光学背景干扰即可检测。首先,我们优化了硝酸纤维素膜上的点印迹免疫分析方法,用于人血清白蛋白(HSA)的定量分析,检测限(LOD)为 0.19ng/mL,信背比(/)为 722。商用量子点被用作参考标记物,其 LOD 为 4.32ng/mL,/为 3,这清楚地表明了 UCNPs 的优势。此外,通过分析加标尿液样本,证实了基于 UCNP 的点印迹分析用于实际样品分析的潜力,检测限(LOD)达到 0.24ng/mL,回收率为 79%至 123%。此外,我们通过将其适应于检测另外两种临床相关生物标志物,前列腺特异性抗原(PSA)和心肌肌钙蛋白(cTn),证明了该检测方法的多功能性和稳健性,在加标血清中的检测限(LOD)分别为 9.4pg/mL 和 0.62ng/mL。最后,分析了用于前列腺癌检查的患者的临床样本,与参考电化学发光免疫分析(回收率为 89%至 117%)具有很强的相关性。研究结果表明,UCNPs 是一种高灵敏度的标记物,可用于开发用于低丰度生物标志物定量分析的点印迹免疫分析方法。