适用于离心分析仪的杜邦茶碱测定法的评估。

Evaluation of the Du Pont Theophylline Assay adapted to a centrifugal analyzer.

作者信息

Boeckx R L, Munson S T

出版信息

Ther Drug Monit. 1985;7(1):95-100. doi: 10.1097/00007691-198503000-00017.

Abstract

We evaluated the Du Pont Theophylline Assay, a particle-enhanced turbidimetric inhibition immunoassay (PETINIA) for the measurement of theophylline in human serum. The procedure was applied to the Cobas-Bio centrifugal analyzer, and was compared to an enzyme immunoassay (EMIT) method and a high performance liquid chromatographic (HPLC) procedure. Day-to-day precision was 3.7% (1 SD coefficient of variation) at 7.1 mg/L, 3.3% at 17.7 mg/L, and 3.9% at 27.8 mg/L. The assay was linear up to 40 mg/L, and the correlation between the PETINIA, EMIT, and HPLC methods was good [PETINIA/EMIT: y = 0.94x + 0.63, r = 0.98, Syx = 1.13; PETINIA/HPLC: y = 1.00x - 0.89, r = 0.99, Syx = 0.66, where Syx is the standard deviation of the residual error of regression], when evaluated using specimens from 135 patients receiving theophylline. No interference from hemolysis, lipemia, icterus, or other methylxanthines was observed. Of the major metabolites of theophylline, only 1,3-dimethyluric acid showed any significant cross-reactivity (2.1 mg/L apparent theophylline at 20 mg/L 1,3-dimethyluric acid). The method is reliable and cost-effective for the measurement of theophylline in serum.

摘要

我们评估了用于测定人血清中茶碱的杜邦茶碱检测法,一种颗粒增强比浊抑制免疫分析法(PETINIA)。该检测程序应用于Cobas - Bio离心分析仪,并与酶免疫分析法(EMIT)和高效液相色谱法(HPLC)进行比较。在7.1 mg/L时,日间精密度为3.7%(变异系数1 SD),在17.7 mg/L时为3.3%,在27.8 mg/L时为3.9%。该检测法在高达40 mg/L时呈线性,当使用135例接受茶碱治疗患者的样本进行评估时,PETINIA法、EMIT法和HPLC法之间的相关性良好[PETINIA/EMIT:y = 0.94x + 0.63,r = 0.98,Syx = 1.13;PETINIA/HPLC:y = 1.00x - 0.89,r = 0.99,Syx = 0.66,其中Syx是回归剩余误差的标准差]。未观察到溶血、脂血、黄疸或其他甲基黄嘌呤的干扰。在茶碱的主要代谢产物中,只有1,3 - 二甲基尿酸显示出任何显著的交叉反应性(在20 mg/L 1,3 - 二甲基尿酸时,表观茶碱为2.1 mg/L)。该方法对于测定血清中的茶碱可靠且具有成本效益。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验