Fetal Diagn Ther. 2024;51(5):432-444. doi: 10.1159/000539271. Epub 2024 Jun 20.
Three-dimensional fractional moving blood volume (3D-FMBV) may provide superior noninvasive measurement of feto-placental perfusion compared to current methods. This study investigated the feasibility and repeatability of producing 3D-FMBV measurements of the placenta, fetal liver, kidney, and brain in a single ultrasound consultation.
The placenta, fetal liver, kidney, and brain were scanned in triplicate using 3D power Doppler ultrasound (3D-PDU) in 48 women ≥22 weeks of gestation with healthy fetuses. 3D-FMBV was calculated by two analyzers. Feasibility was assessed as the percentage of cases where 3D-FMBV could be evaluated; repeatability (intraobserver and interobserver) using two-way mixed measure intraclass correlation coefficients (ICCs).
3D-FMBV was calculated for 100% of scanned organs. Intraobserver ICCs (95% CI) were good to excellent; 0.93 (0.88-0.96) and 0.87 (0.78-0.92) for placenta, 0.95 (0.92-0.97) and 0.98 (0.96-0.99) for fetal liver, 0.96 (0.94-0.98) and 0.91 (0.85-0.95) for fetal kidney, and 0.98 (0.97-0.99) and 0.97 (0.95-0.98) for fetal brain. Interobserver ICCs (95% CI) were 0.50 (0.08-0.73), 0.92 (0.85-0.96), 0.89 (0.78-0.94), and 0.71 (0.46-0.85) for placenta, fetal liver, kidney, and brain.
Feto-placental perfusion assessment with 3D-FMBV is highly reliable in healthy pregnancies ≥22 weeks of gestation and can be feasibly calculated in four feto-placental vascular beds in a single ultrasound consultation.
与当前方法相比,三维分数运动血容量(3D-FMBV)可能提供更优的胎儿胎盘灌注的无创测量。本研究旨在调查在单次超声检查中对胎盘、胎儿肝、肾和脑进行 3D-FMBV 测量的可行性和可重复性。
对 48 例 22 周以上、胎儿健康的孕妇进行三次三维能量多普勒超声(3D-PDU)扫描,计算 3D-FMBV。两名分析员计算 3D-FMBV。以可评估的病例百分比评估可行性;使用双向混合测量组内相关系数(ICC)评估观察者内和观察者间的可重复性。
可计算 100%扫描器官的 3D-FMBV。观察者内 ICC(95%CI)为良好至优秀;胎盘为 0.93(0.88-0.96)和 0.87(0.78-0.92),胎儿肝为 0.95(0.92-0.97)和 0.98(0.96-0.99),胎儿肾为 0.96(0.94-0.98)和 0.91(0.85-0.95),胎儿脑为 0.98(0.97-0.99)和 0.97(0.95-0.98)。观察者间 ICC(95%CI)为 0.50(0.08-0.73)、0.92(0.85-0.96)、0.89(0.78-0.94)和 0.71(0.46-0.85),分别用于胎盘、胎儿肝、肾和脑。
22 周以上健康妊娠中,使用 3D-FMBV 评估胎儿胎盘灌注具有高度可靠性,并且可以在单次超声检查中对四个胎儿胎盘血管床进行可行性计算。