Suppr超能文献

陶瓷动脉瘤夹,带钛制弹簧。

Ceramic Aneurysm Clip with Titanium Spring.

机构信息

Department of Neurosurgery, School of Medicine, Kyungpook National University, Daegu, Republic of Korea.

Department of Neurosurgery, School of Medicine, Kyungpook National University, Daegu, Republic of Korea.

出版信息

World Neurosurg. 2024 Sep;189:256-263. doi: 10.1016/j.wneu.2024.06.105. Epub 2024 Jun 26.

Abstract

OBJECTIVE

To design and evaluate ceramic aneurysm clips with integrated titanium springs, focusing on ergonomic application and precision in neurosurgical procedures.

METHODS

The clip design was executed with precision using Creo Parametric 3D CAD software. It comprises a zirconia body and a titanium spring for durability and consistent tension and features a four-coil hairpin titanium spring for enhanced closing force and a ball-type head for versatile maneuverability during surgery. To assess durability, closing forces were rigorously measured using a force gauge system, comparing the ceramic clip with the standard Mizuho permanent clip over 30 open-close cycles. For the assessment of magnetic resonance (MR) artifacts, both the ceramic and Yasargil clips were evaluated using a 3 Tesla (T) MRI scanner, employing specific imaging sequences.

RESULTS

The straight type ceramic clip's initial closing force was 1.70 N, dropping to 1.22 N after 30 cycles, indicating a retention of 72% of its initial force. In MRI, the ceramic clip displayed significantly lower measurement discrepancies compared to the titanium alloy Yasargil clip, particularly in high-resolution T1-weighted images. The lowest variance was at measurement point L2, where the ceramic clip showed a 3% discrepancy. Furthermore, the ceramic clip yielded clearer images than the titanium alloy clip, particularly at the clip's end.

CONCLUSIONS

Ceramic clips with titanium springs demonstrated satisfactory closing force and superior MRI compatibility, promising enhancements in surgical application and postoperative assessment.

摘要

目的

设计并评估具有集成钛弹簧的陶瓷动脉瘤夹,重点关注神经外科手术中的人体工程学应用和精度。

方法

使用 Creo Parametric 3D CAD 软件精确执行夹的设计。它由氧化锆体和钛弹簧组成,具有耐用性和一致的张力,采用四匝发夹式钛弹簧,可增强闭合力,球形头部可在手术中提供多功能的操纵性。为了评估耐用性,使用力计系统严格测量闭合力,在 30 次打开-关闭循环中比较陶瓷夹与标准 Mizuho 永久夹。为了评估磁共振(MR)伪影,使用特定的成像序列,在 3 特斯拉(T)MRI 扫描仪上评估陶瓷夹和 Yasargil 夹。

结果

直型陶瓷夹的初始闭合力为 1.70N,经过 30 个循环后降至 1.22N,表明其初始力保留了 72%。在 MRI 中,陶瓷夹与钛合金 Yasargil 夹相比,显示出明显更低的测量差异,特别是在高分辨率 T1 加权图像中。在测量点 L2 处,差异最小为 3%。此外,陶瓷夹产生的图像比钛合金夹更清晰,尤其是在夹的末端。

结论

具有钛弹簧的陶瓷夹表现出令人满意的闭合力和优异的 MRI 兼容性,有望增强手术应用和术后评估。