Unexpected Artifact on Tc-Methylene Diphosphonate Bone Scintigraphy.

作者信息

Salih Suliman G, Osman Abdelbagi O, Alkatheeri Ajnas M, Hassaneen Mohamed B

机构信息

Department of Radiography and Medical Imaging, Fatima College of Health Sciences, Al Ain, United Arab Emirates; and

Department of Nuclear Medicine, National Cancer Institute, University of Gezira, Sudan.

出版信息

J Nucl Med Technol. 2024 Dec 4;52(4):360-361. doi: 10.2967/jnmt.124.267795.

Abstract

Asymmetric hot spots in the axial skeleton on bone scintigraphy may confound diagnosis. We describe an unexpected artifact of Tc-methylene diphosphonate near the breast in a 55-y-old woman with breast cancer. The initial whole-body bone scintigraphy revealed a solitary focal lesion in the anterior ribs on the left side. After careful tracking, we determined that this hot spot originated from the adhesive bandage. The patient had placed it in her left front pocket after removing it from the injection site. She was rescanned after the bandage had been removed from her pocket.