Accuracy of a new photometric jaw tracking system in the frontal plane at different recording distances: An in-vitro study.

作者信息

Valenti Chiara, Massironi Domenico, Truffarelli Tiberio, Grande Francesco, Catapano Santo, Eramo Stefano, Tribbiani Giulio, Pagano Stefano

机构信息

PhD student, CISAS "Giuseppe Colombo", University of Padua, Via Venezia, 15, 35131, Padua, Italy; Resident Student, Department of Medicine and Surgery, Faculty of Dentistry, University of Perugia, S. Andrea delle Fratte, 06156 Perugia, Italy.

Private Practice, Via L. Cadorna, 21, 20077 Melegnano, Italy.

出版信息

J Dent. 2024 Sep;148:105245. doi: 10.1016/j.jdent.2024.105245. Epub 2024 Jul 19.

Abstract

OBJECTIVES

To evaluate the accuracy of a new photometric jaw tracking system (JTS) in recording linear vertical movements in the frontal plane at different distances.

METHODS

A mandibular plaster cast of a patient was placed on a simulation machine capable of linear movements along two spatial axes. Cyclops JTS (Itaka) was adapted to the plaster cast, while the head frame was attached to the simulation machine. The latter performed five linear movements from 20 to 40 mm in the y-axis; each movement was repeated five times at five different recording distance (380 to 420 mm). The recorded movements were measured and compared with those obtained with a laser Doppler vibrometer (LDV) for accuracy analysis. Data were statistically processed (α = 0.05).

RESULTS

No statistically significant differences were found between Cyclops and LDV measurements on the y- and z-axes (p = 0.5). Changes in linear vertical motion and distance positions did not affect the accuracy, which remained relatively constant with similar trends and values less than 1 % for each parameter variation. The best condition observed was linear vertical movement of 30 mm at 420 mm (0.010 ± 0.023 mm).

CONCLUSIONS

Cyclops has proven to be an accurate JTS in recording linear vertical movements in the frontal plane at different recording distances. For optimal recordings, the scanner should be placed as close as possible to the markers; excessive vertical movements decreased the accuracy. However, this study has limitations and requires in-vivo confirmations.

CLINICAL SIGNIFICANCE

The tested JTS proved accurate in recording linear vertical movements in the frontal plane. However, given the limitations of the study, further investigation under real conditions is needed to support prosthetic and gnathological rehabilitations.

摘要

目的

评估一种新型光度式颌骨跟踪系统(JTS)在记录不同距离下额平面线性垂直运动时的准确性。

方法

将一名患者的下颌石膏模型放置在一台能够沿两个空间轴进行线性运动的模拟机器上。将Cyclops JTS(伊塔卡)适配到石膏模型上,同时将头架连接到模拟机器上。模拟机器在y轴上进行了五次20至40毫米的线性运动;每次运动在五个不同的记录距离(380至420毫米)下重复五次。测量记录的运动,并与使用激光多普勒振动计(LDV)获得的运动进行比较,以进行准确性分析。对数据进行统计学处理(α = 0.05)。

结果

在y轴和z轴上,Cyclops测量值与LDV测量值之间未发现统计学上的显著差异(p = 0.5)。线性垂直运动和距离位置的变化不影响准确性,每个参数变化时准确性相对保持恒定,趋势相似且值小于1%。观察到的最佳条件是在420毫米处进行30毫米的线性垂直运动(0.010±0.023毫米)。

结论

已证明Cyclops是一种准确的JTS,可用于记录不同记录距离下额平面的线性垂直运动。为了获得最佳记录,扫描仪应尽可能靠近标记放置;过度的垂直运动会降低准确性。然而,本研究存在局限性,需要进行体内验证。

临床意义

经测试的JTS在记录额平面线性垂直运动方面被证明是准确的。然而,鉴于本研究的局限性,需要在实际条件下进行进一步研究,以支持假体修复和口腔颌面康复。