一项比较研究:即时检验以及采用不同生物基质和抗凝剂的中心实验室分析仪测定肌钙蛋白I时可能存在的偏差

A Comparison Study: Possible Bias in Troponin I Measurement Obtained with a Point of Care Testing and a Central Laboratory Analyzers Employing Different Biological Matrices and Anticoagulants.

作者信息

Sammartano Antonino, Buonocore Ruggero, Fiorini Roberto, Dieci Elisabetta, Di Franco Anna, Di Stasi Bruna, Tortorella Giovanni, Ippolito Luigi

机构信息

U.O. Clinical Pathology, Medical and Diagnostics Department P.O. Fidenza, Azienda AUSL of Parma, 43125 Parma, Italy.

U.O. Biochemical, Department of Clinical Patology, Ospedale Guglielmo da Saliceto of Piacenza, 29121 Piacenza, Italy.

出版信息

Diagnostics (Basel). 2024 Jul 10;14(14):1482. doi: 10.3390/diagnostics14141482.

Abstract

BACKGROUND

high-sensitive cardiac TroponinI (hs-cTnI) is widely used for diagnosis of acute coronary syndromes. The latest recommendation for hs-cTnI determination is the protocol 0-1 h finalized to improve the rule out accuracy of the test. A Point of Care Testing able to guarantee these performances could be very useful due to reducing the turnaround time and ruling out patients suspected of ACS, especially by using biological matrices that are not required for centrifuge. The aim of our work is to compare the results for hs-cTnI obtained using different biological matrices and anticoagulants, obtained between Atellica VTLi hs-cTnI POCT and Access AccuTnI+3 DxI800 performances, in order to establish a possible bias derived directly from these pre-analytical conditions.

METHODS

Li-heparinized pool samples were primary employ for hs-cTnI with Atellica VTLi as whole blood, then centrifuged and tested on Atellica VTLi and DxI800. KEDTA pool samples were centrifuged and measured on DxI800 too. A comparison of methods was performed according to CLSI_EP-09A2 protocol. Constant and proportional errors were investigated with Deming regression. Bias between methods was evaluated with the Bland Altman test.

RESULTS

comparing whole blood lithium heparin results obtained with Atellica versus lithium heparin and KEDTA plasma tested on DxI 800, the Deming regression revealed a proportional error, whereas in both cases Bland Altman highlighted a minimal underestimation. A similar performance was revealed when considering plasma lithium heparin tested on Atellica versus lithium heparin and KEDTA plasma obtained with DxI800, confirming the same underestimation. Considering values close to the cut off, no significant differences were found.

CONCLUSIONS

in the laboratory, the estimation of the bias of two different analyzers is pivotal. Once more this is crucial when different biological matrices and anticoagulants are employed for the analysis. Our study demonstrates that no significant differences among the two matrices are present when comparing Atellica and DxI800 performances.

摘要

背景

高敏心肌肌钙蛋白I(hs-cTnI)广泛用于急性冠状动脉综合征的诊断。hs-cTnI测定的最新建议是0-1小时方案,其最终确定是为了提高检测的排除准确性。一种能够保证这些性能的即时检测(POCT)由于可以缩短周转时间并排除疑似急性冠状动脉综合征(ACS)的患者,特别是通过使用无需离心的生物基质,可能会非常有用。我们工作的目的是比较使用不同生物基质和抗凝剂获得的hs-cTnI结果,这些结果是在Atellica VTLi hs-cTnI POCT和Access AccuTnI+3 DxI800性能之间获得的,以便确定直接源自这些分析前条件的可能偏差。

方法

以肝素锂抗凝的混合样本最初作为全血用于Atellica VTLi检测hs-cTnI,然后离心并在Atellica VTLi和DxI800上进行检测。乙二胺四乙酸钾(KEDTA)混合样本也进行离心并在DxI800上测量。根据CLSI_EP-09A2方案进行方法比较。用Deming回归研究恒定误差和比例误差。用Bland Altman检验评估方法之间的偏差。

结果

将Atellica获得的全血肝素锂结果与在DxI 800上检测的肝素锂和KEDTA血浆结果进行比较,Deming回归显示存在比例误差,而在两种情况下Bland Altman均突出显示了最小程度的低估。当考虑在Atellica上检测的肝素锂血浆与用DxI800获得的肝素锂和KEDTA血浆时,显示出类似的性能,证实了相同程度的低估。考虑接近临界值的值时,未发现显著差异。

结论

在实验室中,估计两种不同分析仪的偏差至关重要。当采用不同的生物基质和抗凝剂进行分析时,这一点再次变得至关重要。我们的研究表明,在比较Atellica和DxI800性能时,两种基质之间不存在显著差异。