Song Hyun Seok, Joh Yongwon, Kim Hyungsuk
Department of Orthopedic Surgery, Eunpyeong St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul 03312, Republic of Korea.
Life (Basel). 2024 Jul 18;14(7):888. doi: 10.3390/life14070888.
The human clavicle's unique S-shaped, three-dimensional structure complicates fracture management. This study evaluated the anatomical conformity of pre-contoured anatomical plates using 3D-printed clavicle models. CT scans from 30 patients (15 males and 15 females) were used to create these models. Three brands of distal clavicle plate systems (Acumed, Synthes, and Arthrex) were tested for fit. Measurements included the distance from the distal end of the clavicle to the plate's lateral end, the gap between the clavicle and the plate, and the overhang distance. Results showed significant differences in clavicle length between sexes, with men having a mean length of 156.1 ± 7.6 mm and women 138.4 ± 4.3 mm, both with normal distribution ( > 0.05). The mean lateral distance was 7.9 ± 1.7 mm, and the mean medial gap was 3.6 ± 3.0 mm, showing no significant differences between products or sexes. The mean overhang distance was 5.8 ± 4.6 mm, with larger values in women for the Acumed ( = 0.037) and Arthrex ( = 0.000) plates. Overall, pre-contoured plates exhibited notable discrepancies, especially in shorter clavicles.
人类锁骨独特的S形三维结构使骨折治疗变得复杂。本研究使用3D打印的锁骨模型评估了预塑形解剖钢板的解剖学贴合度。利用30例患者(15例男性和15例女性)的CT扫描创建这些模型。对三个品牌的锁骨远端钢板系统(Acumed、Synthes和Arthrex)进行了贴合度测试。测量内容包括从锁骨远端到钢板外侧端的距离、锁骨与钢板之间的间隙以及悬垂距离。结果显示,两性之间锁骨长度存在显著差异,男性平均长度为156.1±7.6毫米,女性为138.4±4.3毫米,均呈正态分布(>0.05)。平均外侧距离为7.9±1.7毫米,平均内侧间隙为3.6±3.0毫米,产品或性别之间无显著差异。平均悬垂距离为5.8±4.6毫米,Acumed(P = 0.037)和Arthrex(P = 0.000)钢板在女性中的值更大。总体而言,预塑形钢板存在显著差异,尤其是在较短的锁骨中。