Wagner Marcus, Stoffel-Wagner Birgit, Zur Berndt
Institute of Clinical Chemistry and Clinical Pharmacology, University Hospital Bonn, Bonn, Germany.
Institute for Experimental Haematology and Transfusion Medicine, University Hospital Bonn, Bonn, Germany.
Hemoglobin. 2024 May;48(3):209-211. doi: 10.1080/03630269.2024.2360450. Epub 2024 Aug 5.
Hemoglobin Strasbourg is a rare high oxygen affinity hemoglobin variant which leads to secondary erythrocytosis. This variant is caused by a gene mutation c.71T > A resulting in an amino acid exchange on position 23 of the β globin chain (p.Val23Asp.). The influence of Hb Strasbourg on HbA1c measurement has not been studied to date. For patients with hemoglobin variants it is important to know whether possible interferences exist with the measurement of HbA1c. We therefore investigated the influence of Hb Strasbourg on HbA1c measurement with two different HPLC (high-performance liquid chromatography) systems and one turbidimetric immunoassay in two non-diabetic brothers who are heterozygous carriers of Hb Strasbourg. The examined tests are all used in routine diagnostics. In the case of Hb Strasbourg, the HbA1c measured by HPLC showed lower results than those obtained by the immunoassay. We conclude that HbA1c is underestimated when measured with these methods as glycated Hb Strasbourg is most likely not co-eluting with HbA1c in HPLC.
斯特拉斯堡血红蛋白是一种罕见的高氧亲和力血红蛋白变体,可导致继发性红细胞增多症。该变体由基因突变c.71T > A引起,导致β珠蛋白链第23位的氨基酸交换(p.Val23Asp)。迄今为止,尚未研究过斯特拉斯堡血红蛋白对糖化血红蛋白(HbA1c)测量的影响。对于患有血红蛋白变体的患者,了解HbA1c测量是否可能存在干扰非常重要。因此,我们在两名携带斯特拉斯堡血红蛋白的非糖尿病兄弟中,使用两种不同的高效液相色谱(HPLC)系统和一种比浊免疫测定法,研究了斯特拉斯堡血红蛋白对HbA1c测量的影响。所检测的方法均用于常规诊断。对于斯特拉斯堡血红蛋白,HPLC测量的HbA1c结果低于免疫测定法获得的结果。我们得出结论,使用这些方法测量时,HbA1c被低估,因为在HPLC中,糖化的斯特拉斯堡血红蛋白很可能不会与HbA1c共洗脱。