一种自动血液培养系统:使用马尔萨斯微生物生长分析仪检测厌氧菌。

An automated blood culture system: the detection of anaerobic bacteria using a Malthus Microbiological Growth Analyser.

作者信息

McMaster J P, Barr J G, Campbell R R, Bennett R B, Smyth E T

出版信息

Ulster Med J. 1985 Oct;54(2):133-9.

Abstract

The Malthus Microbiological Growth Analyser has proved to be sensitive in detecting conductivity changes due to anaerobic metabolism in a number of widely used blood culture media. Freshly prepared cooked meat media and Thiol medium yielded the greatest gross conductivity changes, and were more sensitive of anaerobic metabolism than other media. Failure of the instrument to detect anaerobic metabolism was a problem particularly associated with growth in the thioglycollate medium. False positive detections of growth were attributed to a number of factors including electrode instability (6.0%) and bacterial contamination (8.75%).

摘要

马尔萨斯微生物生长分析仪已被证明在检测多种广泛使用的血培养介质中因厌氧代谢引起的电导率变化方面很灵敏。新制备的熟肉培养基和硫醇培养基产生的总电导率变化最大,并且比其他培养基对厌氧代谢更敏感。该仪器未能检测到厌氧代谢是一个特别与硫乙醇酸盐培养基中的生长相关的问题。生长的假阳性检测归因于多种因素,包括电极不稳定(6.0%)和细菌污染(8.75%)。

相似文献

7
Electrode and electrolyte impedance in the detection of bacterial growth.
IEEE Trans Biomed Eng. 1981 May;28(5):403-10. doi: 10.1109/TBME.1981.324723.

本文引用的文献

1
A cooked meat blood culture medium; shelf-life and clinical evaluation compared with other systems.
J Infect. 1980 Sep;2(3):247-58. doi: 10.1016/s0163-4453(80)90698-2.
4
Blood culture technique based on centrifugation: clinical evaluation.
J Clin Microbiol. 1976 Mar;3(3):258-63. doi: 10.1128/jcm.3.3.258-263.1976.
6
Microcalorimetry as a tool for evaluation of blood culture media.
J Clin Microbiol. 1977 Apr;5(4):393-6. doi: 10.1128/jcm.5.4.393-396.1977.