Suppr超能文献

雷击伤后起搏器完全失效 1 例报告。

Complete pacemaker failure following lightning strike injury: A case report.

机构信息

Department of Cardiology, Hospital Universitário Onofre Lopes, Universidade Federal do Rio Grande do Norte, Natal, Brazil.

出版信息

Pacing Clin Electrophysiol. 2024 Nov;47(11):1503-1506. doi: 10.1111/pace.15064. Epub 2024 Aug 19.

Abstract

INTRODUCTION

Modern cardiovascular implantable electronic devices (CIEDs) have mechanisms that prevent damage from external electric shocks, and malfunction following accidental electrocution is rare. However, the effects of lightning injuries in patients with CIEDs are uncertain.

CASE PRESENTATION

A 74-year-old man with a dual-chamber pacemaker due to complete heart block was struck by a lightning while farming. He had no serious injury at the time and sought medical evaluation 1 month later, when he presented with asymptomatic bradycardia. Device interrogation suggested major battery and lead damage, requiring extraction and subsequent placement of a new pacing system.

DISCUSSION

While a previous report depicted pacing threshold elevation without extensive device impairment, our patient presented with major damage to the whole pacing system. The factors contributing to these divergent outcomes are unclear. Differences in injury mechanism, pacemaker model, and the pattern of electric current dispersion within the device may each play a part in this discrepancy.

摘要

简介

现代心血管植入式电子设备(CIEDs)具有防止外部电击损坏的机制,意外触电后发生故障的情况很少见。然而,CIED 患者遭受雷击的影响尚不确定。

病例介绍

一名 74 岁男性因完全性心脏传导阻滞植入双腔起搏器,在务农时被闪电击中。当时他没有受重伤,一个月后因无症状心动过缓就诊并接受了医疗评估。设备检测提示电池和导联严重损坏,需要取出并随后植入新的起搏系统。

讨论

尽管之前有报道显示起搏阈值升高但设备无广泛损害,但我们的患者整个起搏系统都受到严重损坏。导致这些不同结果的因素尚不清楚。损伤机制、起搏器型号以及电流在设备内的分布模式的差异都可能在这种差异中起作用。

文献AI研究员