Atmospheric electrical discharge is an extremely powerful natural phenomenon which can have dangerous and lethal effects on the human body. However, there is no evidence to indicate whether and, if so, to what extent the electric current travelling through the body can affect proper pacemaker function.
An 80-year-old patient admitted to emergency department after being struck by a lightning bolt while riding a bike. The patient had a DDD pacemaker implanted 4 years prior to the incident. The ECG on admission depicted pacemaker spikes and native sinus rhythm at 50-60 b.p.m. On the 3rd day after admission the patient developed recurrent pacing-induced tachycardia. Pacemaker interrogation showed high pacing thresholds (failure to pace in the atrial channel). When the patient's condition stabilized she was transferred to the tertiary hospital for transcutaneous lead extraction. The extracted pacing system was sent to Biotronik for thorough evaluation.
Injuries due to a lightning strike are considered a rare occurrence but being struck by lightning with a pacemaker or an ICD is even less common. In the present case, the cause of cardiac arrhythmia was most probably electrical burn at the endocardial-electrode interface and a sudden elevation of the pacing threshold leading to transient pacing failure in both PM channels. To the best of our knowledge, in this case presentation we first described permanent lightning-induced pacemaker dysfunction.
大气放电是一种极其强大的自然现象,可对人体产生危险和致命影响。然而,尚无证据表明流经人体的电流是否以及在何种程度上会影响起搏器的正常功能。
一名80岁患者在骑自行车时遭雷击后被送往急诊科。该患者在事件发生前4年植入了DDD起搏器。入院时的心电图显示起搏器脉冲和50 - 60次/分钟的窦性心律。入院后第3天,患者出现反复的起搏诱发心动过速。起搏器程控显示起搏阈值升高(心房通道起搏失败)。患者病情稳定后被转至三级医院进行经皮导线拔除。拔除的起搏系统被送往百多力公司进行全面评估。
雷击造成的伤害被认为是罕见的,但带着起搏器或植入式心律转复除颤器(ICD)遭雷击则更为罕见。在本病例中,心律失常的原因很可能是心内膜 - 电极界面的电灼伤以及起搏阈值突然升高,导致双腔起搏器通道出现短暂起搏失败。据我们所知,在本病例报告中,我们首次描述了永久性雷击诱发的起搏器功能障碍。