This paper proposes measurement and reduction of eye dose in real time for the physician and the assistant performing fluoroscopy guided arterial puncture. Eye dose rates were measured for 30 fluoroscopy-guided punctures of bilateral femoral arteries in pigs. Fifteen fluoroscopy-guided punctures were performed using real time radiation dosimeter without auditory and visual feedback and other fifteen punctures were done using real time radiation dosimeter with visual and auditory feedback worn on forehead by an interventional cardiologist having experience of more than 10 years. The mean radiation dose rate for eyes of physician during arterial puncturing with real time radiation dosimeter with auditory feedback was 0.07 mSv/h (n = 15) whereas it was 0.18 mSv/h (n = 15) without visual and auditory feedback. The percentage of reduction with the device was 61% for eyes. In case of assistant the reduction was 33% for eyes (n = 15). The real time visual and auditory feedback dosimeter has reduced the eye dose rate of the physician and assistant and also helped him staying away from the X-ray source. Real time radiation dosimeters can be an effective tool to measure and reduce the dose to the eyes. The radiation eye dose rate for physician and assistant was significantly reduced by using real time radiation dosimeter with visual and auditory feedback. The real time radiation dosimeter not only helps in measuring but also help in minimizing the radiation dose rate for the physician and assistant in real time.
本文提出了一种实时测量和减少医生和助手在进行透视引导动脉穿刺时眼部剂量的方法。对 30 例猪双侧股动脉透视引导穿刺进行了眼部剂量率测量。15 例透视引导穿刺使用实时辐射剂量计进行,无听觉和视觉反馈,另外 15 例穿刺使用实时辐射剂量计进行,由一位具有 10 年以上经验的介入心脏病专家在前额佩戴视觉和听觉反馈。有听觉反馈的实时辐射剂量计测量到的医生眼部辐射剂量率为 0.07mSv/h(n=15),而无视觉和听觉反馈时为 0.18mSv/h(n=15)。该设备使眼部剂量降低了 61%。对于助手,眼部剂量降低了 33%(n=15)。实时视觉和听觉反馈剂量计降低了医生和助手的眼部剂量率,并帮助他们远离 X 射线源。实时辐射剂量计是一种测量和减少眼部剂量的有效工具。使用带视觉和听觉反馈的实时辐射剂量计,医生和助手的眼部辐射剂量率显著降低。实时辐射剂量计不仅有助于测量,还可以帮助医生和助手实时最小化辐射剂量率。