Rassow J
Strahlentherapie. 1985 Feb;161(2):96-7.
The LET dependence of the response of different glow-peaks of CaF2 : Tm (TLD-300) detectors can be used for separate simultaneous determination of neutron and gamma-ray absorbed doses in mixed neutron-photon radiation fields. The principle the pre-conditions of this "two-peak TLD method" are explained using experimental results with d(14) + Be neutrons (EN = 5,8 MeV).
CaF2 : Tm(TLD - 300)探测器不同发光峰响应的传能线密度(LET)依赖性可用于在混合中子 - 光子辐射场中分别同时测定中子和γ射线吸收剂量。利用d(14)+ Be中子(EN = 5.8 MeV)的实验结果解释了这种“双峰热释光剂量计方法”的原理和前提条件。