Sphenoidal electrode positions and basal EEG during long term monitoring.

作者信息

Wilkus R J, Thompson P M

出版信息

Epilepsia. 1985 Mar-Apr;26(2):137-42. doi: 10.1111/j.1528-1157.1985.tb05397.x.

Abstract

In 33 patients undergoing extended EEG monitoring, stereoscopic basal skull radiographs demonstrated the positions of Teflon-coated, 40-gauge, multistranded, silver wire electrodes shortly after their insertion and just before their removal an average of 5.12 days later. Although the electrodes had been firmly secured to the skin with plastic surgical adhesive tape and there was no evidence that they had been manually pulled out, many wee kinked and relatively superficially situated at the end, compared with the onset, of each monitoring period. Electrodes moved similar amounts bilaterally and migrated further as the days passed. Sphenoidal wire tips that had been an average of 7.02 mm lateral to the foramen ovale when placed eventually averaged 15.79 mm from this landmark. There was no relationship between distance moved by the electrodes and sex, age, or number or type of seizures exhibited. Review of EEG records of interictal discharges and seizures revealed no differences attributable to the relative electrode positions.

摘要

在33例接受长时间脑电图监测的患者中,立体基底颅骨X线片显示了涂有特氟龙的40号多股银丝电极在插入后不久以及平均5.12天后即将取出前的位置。尽管电极已用外科手术塑料胶带牢固地固定在皮肤上,且没有证据表明它们是被手动拔出的,但与每个监测期开始时相比,许多电极在末期出现了扭结且位置相对表浅。电极双侧移动的幅度相似,且随着时间推移移动得更远。放置时距卵圆孔平均外侧7.02毫米的蝶骨电极尖端最终距该标志平均为15.79毫米。电极移动的距离与性别、年龄或癫痫发作的次数及类型之间没有关系。对发作间期放电和癫痫发作的脑电图记录进行回顾,未发现因电极相对位置不同而产生的差异。