The intrathecal baclofen pump is an effective treatment option for patients with severe spasticity. In children, subfascial pump placement is often preferred given concerns for infection and wound healing. However, this approach is not without risk, and rare complications, such as peritoneal pump migration, can occur.
The authors describe three pediatric cases of peritoneal pump migration at their institution over the past 14 years (3/545, 0.5%). All three patients had low body weight (below the 39th percentile), and two had scoliosis requiring surgery. All pumps had been placed using the subfascial technique. The first case occurred 6 months postplacement, and the pump was not replaced. Cases 2 and 3 occurred at 2 and 3 years postplacement, respectively, and both pumps were replaced.
The authors conclude that peritoneal pump migration, although uncommon, can occur in patients with subfascial pump placement, and providers should have a low threshold of suspicion for repeat imaging prior to refilling if the pump's location has migrated. Potential contributing factors to pump migration include a patient's small size, a larger pump size (40-mL pump), and scoliosis. All these factors should be considered during pump placement, and surgeons can consider using a 20-ml pump for smaller patients. https://thejns.org/doi/abs/10.3171/CASE24290.
鞘内注射巴氯芬泵是治疗严重痉挛患者的一种有效选择。对于儿童患者,考虑到感染和伤口愈合问题,通常更倾向于采用筋膜下泵植入术。然而,这种方法并非没有风险,可能会出现罕见的并发症,如腹膜泵移位。
作者描述了过去14年中在他们机构发生的3例小儿腹膜泵移位病例(545例中有3例,0.5%)。所有3例患者体重均较低(低于第39百分位数),其中2例患有脊柱侧弯需要手术治疗。所有泵均采用筋膜下技术植入。第1例在植入后6个月发生,泵未更换。第2例和第3例分别在植入后2年和3年发生,两个泵均进行了更换。
作者得出结论,腹膜泵移位虽然不常见,但在采用筋膜下泵植入的患者中可能会发生,如果泵的位置发生移位,在重新注药前,医护人员对重复成像检查的怀疑阈值应较低。泵移位的潜在促成因素包括患者体型小、泵的尺寸较大(40毫升泵)和脊柱侧弯。在泵植入过程中应考虑所有这些因素,对于体型较小的患者,外科医生可考虑使用20毫升的泵。https://thejns.org/doi/abs/10.3171/CASE24290