Suppr超能文献

[一种新型膈肌活动度测量方法的验证研究:在肝肾/脾肾切面使用相控阵探头评估膈肌活动度]

[Validation study of a novel diaphragmatic excursion measurement method: evaluation of diaphragmatic excursion by phased-array probe in hepato-renal/spleno-renal section].

作者信息

Ma Junyu, Zhai Shanshan, Sun Xiaocong, Li Chen, Duan Jun

机构信息

Department of Critical Care Medicine, China-Japan Friendship Hospital, Beijing 100029, China. Corresponding author: Duan Jun, Email:

出版信息

Zhonghua Wei Zhong Bing Ji Jiu Yi Xue. 2024 Jul;36(7):740-744. doi: 10.3760/cma.j.cn121430-20240207-00118.

Abstract

OBJECTIVE

To investigate the accuracy and reproducibility of diaphragmatic excursion (DE) measurements through hepato-renal/spleno-renal section as a novel method for assessing diaphragmatic function.

METHODS

Twelve healthy participants were recruited. Each participant underwent DE measurements performed by four operators with varying levels of experience using traditional methods (liver/spleen section) and novel methods (hepato-renal/spleno-renal section), respectively. Among them, two experienced operators were critical care clinicians, and diaphragmatic ultrasound was performed in more than 50 cases. The other two inexperienced operators were respiratory therapists, with less than 10 cases of diaphragmatic ultrasound operations, who received a 2-hour theoretical and operational training before the study. Operators initially used the conventional method with a 1.5-6.0 MHz convex probe in M-mode, placing the sampling line perpendicular to the diaphragm at the point of maximum excursion, and the liver/spleen section DE was determined during normal breathing of participant. Then, they used the novel method with a 1.6-4.5 MHz phased array probe to observe diaphragmatic movement cranio-caudally along the mid-axillary line, employing anatomic M-mode with the sampling line placed perpendicular to the diaphragm at the level of the renal midpoint, and the DE of the hepato-renal/spleno-renal section was measured during normal breathing. The liver and hepato-renal sections were used to assess the right diaphragm, and spleen and spleno-renal sections were used to assess the left diaphragm. Correlation analysis of DE measurements from different sections was conducted using the Deming method, while consistency was assessed using the Bland-Altman method. The consistency of clinical acceptability was defined as the absence of fixed and proportional bias, with a difference of two standard deviations less than 40% of the mean measurement value. Percentage consistency limit = two standard deviations of the differences between measurements/mean measurement value×100%.

RESULTS

Four operators performed image scans of DE in all four sections for each of the twelve subjects, with a high DE acquisition rate of 100% (48/48) for hepato-renal and spleno-renal sections, followed by the liver section [91.7% (44/48)] and the spleen section [66.7% (32/48)], particularly for the left diaphragm assessment, where the DE acquisition rate of spleno-renal section was significantly higher than that of traditional spleen section (P < 0.01). The overall measurement results showed that no significant difference was found in DE determined via the hepato-renal and spleno-renal sections using the novel method (cm: 1.64±0.10 vs. 1.55±0.14, P > 0.05), and they were significantly higher than those determined via the conventional liver and spleen sections (cm: hepato-renal section vs. liver section was 1.64±0.10 vs. 1.44±0.09, spleno-renal section vs. spleen section was 1.55±0.14 vs. 1.09±0.14, both P < 0.01). Correlation analysis revealed good correlations of DE between hepato-renal section and spleno-renal section, between liver section and hepato-renal section, between liver section and spleno-renal section (r values were 0.62, 0.59, and 0.42, all P < 0.01). Consistency analysis showed that the consistency in DE between hepato-renal section and spleno-renal section, as well as between liver section and hepato-renal section was good (both % consistency limits < 40%). However, the DE measured in the spleen section were not correlated with the other three sections, and there was no inconsistency (all % consistency limits > 40%). There was no statistically significant difference in DE measured by the four operators in the liver, spleen, hepato-renal, and spleno-renal sections (cm: 1.49±0.34, 1.44±0.37, 1.43±0.30, and 1.40±0.27 in liver section; 1.10±0.36, 1.05±0.18, 1.09±0.22, and 1.06±0.26 in spleen section; 1.67±0.43, 1.57±0.34, 1.63±0.32, and 1.66±0.36 in hepato-renal section; 1.45±0.33, 1.48±0.34, 1.50±0.24, and 1.65±0.26 in spleno-renal section; all P > 0.05). According to the clinically acceptable range of consistency limits, the DE measured by the four operators in all four sections showed good consistency (all % consistency limits < 40%).

CONCLUSIONS

The novel method of measuring DE through hepato-renal/spleno-renal sections is accurate, highly reproducible, and has a high acquisition rate, serving as a viable alternative to the conventional method involving the liver/spleen section.

摘要

目的

探讨通过肝肾/脾肾切面测量膈肌移动度(DE)作为评估膈肌功能新方法的准确性和可重复性。

方法

招募12名健康参与者。每位参与者分别由四名经验水平不同的操作者采用传统方法(肝脏/脾脏切面)和新方法(肝肾/脾肾切面)进行DE测量。其中两名经验丰富的操作者为重症监护临床医生,进行过50余例膈肌超声检查。另外两名经验不足的操作者为呼吸治疗师,膈肌超声操作少于10例,在研究前接受了2小时的理论和操作培训。操作者最初使用1.5 - 6.0 MHz凸阵探头以M型模式采用传统方法,将取样线垂直于膈肌最大移动点放置,在参与者正常呼吸时测定肝脏/脾脏切面的DE。然后,他们使用1.6 - 4.5 MHz相控阵探头采用新方法,沿腋中线从头尾方向观察膈肌运动,采用解剖M型模式,将取样线垂直于肾中点水平的膈肌放置,在正常呼吸时测量肝肾/脾肾切面的DE。肝脏和肝肾切面用于评估右侧膈肌,脾脏和脾肾切面用于评估左侧膈肌。采用戴明方法对不同切面的DE测量进行相关性分析,采用布兰德 - 奥特曼方法评估一致性。临床可接受一致性定义为无固定和比例偏差,差值的两个标准差小于平均测量值的40%。一致性百分比限值 = 测量值差异的两个标准差/平均测量值×100%。

结果

四名操作者对12名受试者的所有四个切面均进行了DE图像扫描,肝肾和脾肾切面的DE采集率高达100%(48/48),其次是肝脏切面[91.7%(44/48)]和脾脏切面[66.7%(32/48)],特别是在评估左侧膈肌时脾肾切面的DE采集率显著高于传统脾脏切面(P < 0.01)。总体测量结果显示,采用新方法通过肝肾和脾肾切面测定的DE无显著差异(cm:1.64±0.10 vs. 1.55±0.14,P > 0.05),且显著高于通过传统肝脏和脾脏切面测定的DE(cm:肝肾切面与肝脏切面为1.64±0.10 vs. 1.44±0.09,脾肾切面与脾脏切面为1.55±0.14 vs. 1.09±0.14,均P < 0.01)。相关性分析显示肝肾切面与脾肾切面、肝脏切面与肝肾切面、肝脏切面与脾肾切面之间的DE具有良好相关性(r值分别为0.62、0.59和0.42,均P < 0.01)。一致性分析表明肝肾切面与脾肾切面之间以及肝脏切面与肝肾切面之间的DE一致性良好(一致性百分比限值均 < 40%)。然而,脾脏切面测量的DE与其他三个切面不相关,且无不一致情况(所有一致性百分比限值 > 40%)。四名操作者在肝脏、脾脏、肝肾和脾肾切面测量的DE无统计学显著差异(cm:肝脏切面分别为1.49±0.34、1.44±0.37、1.43±0.30和1.40±0.27;脾脏切面分别为1.10±0.36、1.05±0.18、1.09±0.22和1.06±0.26;肝肾切面分别为1.67±0.43、1.57±0.34、1.63±0.32和1.66±0.36;脾肾切面分别为1.45±0.33、1.48±0.34、1.50±0.24和1.65±0.26;均P > 0.05)。根据临床可接受的一致性限值范围,四名操作者在所有四个切面测量的DE显示出良好的一致性(所有一致性百分比限值 < 40%)。

结论

通过肝肾/脾肾切面测量DE的新方法准确、高度可重复且采集率高,可作为涉及肝脏/脾脏切面的传统方法的可行替代方法。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验