Suppr超能文献

增强现实辅助下颌弓口内扫描:概念验证性初步临床研究。

Augmented reality assisted intraoral scanning of mandibular arch: A proof-of-concept pilot clinical study.

机构信息

Department of Prosthetic Dental Sciences, College of Dentistry, King Saud University, Riyadh, Saudi Arabia.

Prosthodontics Department, Faculty of Dentistry, Cairo University, Egypt.

出版信息

J Dent. 2024 Nov;150:105351. doi: 10.1016/j.jdent.2024.105351. Epub 2024 Sep 10.

Abstract

OBJECTIVES

To investigate whether the scanning time, trueness and number of photos are influenced when augmented reality (AR) heads-up display (HUD) is utilized during the intraoral scan of fully dentate mandibular arches.

METHODS

A total of 10 patients (6 females and 4 males) were included. The mandibular arch of each patient was scanned twice using an intraoral scanner (Trios4 Pod IOS: 3Shape): one with and one without AR-HUD (ML2; Magic Leap). Further, alginate impression was taken, and the cast was digitized to acquire the reference model for trueness comparison (T310, Medit). The scan time and number of photos were recorded. Trueness was evaluated qualitatively and quantitatively using colored heat maps and RMSE values respectively. t-test was used to evaluate the difference in scan time, trueness, and number of photos between the two groups (α = .05).

RESULTS

AR-assisted IOS resulted in significantly faster scan time (44 s) compared to the time consumed following conventional scan method without AR-HUD (63 s) (P = <0.001). The number of photos was also significantly less with AR-assisted IOS (836) compared to IOS using conventional technique without AR-HUD (1209) P = <0.001. No statistical difference was detected in RMSE between the test groups.

CONCLUSIONS

Integration of AR technology with IOS process represents a promising potential to acquire digital impressions with reduced scan acquisition time and reduced images count while simultaneously maintaining the trueness of the acquired scans.

CLINICAL SIGNIFICANCE

Augmented Reality presents an emerging potential in Prosthodontics to acquire digital impressions with decreased number of images and acquisition time.

摘要

目的

研究在全牙列下颌弓的口内扫描过程中使用增强现实(AR)平视显示器(HUD)时,扫描时间、准确性和照片数量是否会受到影响。

方法

共纳入 10 名患者(6 名女性和 4 名男性)。每位患者的下颌弓均使用口内扫描仪(Trios4 Pod IOS:3Shape)扫描两次:一次使用 AR-HUD(ML2;Magic Leap),一次不使用 AR-HUD(ML2;Magic Leap)。此外,还进行了藻酸盐印模,并对模型进行数字化以获取准确性比较的参考模型(T310,Medit)。记录扫描时间和照片数量。分别使用彩色热图和 RMSE 值定性和定量评估准确性。使用 t 检验评估两组之间扫描时间、准确性和照片数量的差异(α=0.05)。

结果

与不使用 AR-HUD 的传统扫描方法相比,AR 辅助 IOS 扫描时间明显更快(44 s)(P<0.001)。使用 AR 辅助 IOS 的照片数量也明显少于不使用 AR-HUD 的 IOS(836 张与 1209 张)(P<0.001)。测试组之间的 RMSE 无统计学差异。

结论

将 AR 技术与 IOS 流程集成代表了一种很有前途的潜力,可以在保持获取扫描的准确性的同时,减少扫描采集时间和图像数量,从而获得数字化印模。

临床意义

增强现实技术在修复学中具有很大的潜力,可以减少获取数字化印模的图像数量和采集时间。