Suppr超能文献

Combined Lactate Glucose Ratio as a Novel Marker for Rapid Diagnosis of Cerebrospinal Fluid Bacterial Infection in Neurosurgical Patients: Diagnostic Accuracy Study and Benchtop Analyzer Correlation.

作者信息

Soon BeeHong, Jaafar Ainul S, A Bakar Azizi, Narayanan Vairavan

机构信息

Division of Neurosurgery, Department of Surgery, Faculty of Medicine, Universiti Malaya, Kuala Lumpur, Malaysia; Neurosurgery Unit, Department of Surgery, Faculty of Medicine, Universiti Kebangsaan Malaysia, Kuala Lumpur, Malaysia.

Neurosurgery Unit, Department of Surgery, Faculty of Medicine, Universiti Kebangsaan Malaysia, Kuala Lumpur, Malaysia.

出版信息

World Neurosurg. 2024 Nov;191:e607-e621. doi: 10.1016/j.wneu.2024.09.012. Epub 2024 Sep 10.

Abstract

OBJECTIVE

This study aimed to assess the diagnostic accuracy of a novel marker, the combined lactate glucose ratio (CLGR), in identifying cerebrospinal fluid (CSF) bacterial infection (CBI) in neurosurgical patients. Additionally, it seeks to establish cutoff values for CLGR and evaluate the reliability of measurement using blood gas analyzer (BGA).

METHODS

CSF samples were collected from 2 neurosurgical centers in Kuala Lumpur, Malaysia, between January 2022 and October 2023. Conventional markers and CLGR were quantified using standard laboratory methods, with BGA utilized for measurement when feasible. Samples were categorized into confirmed CBI-positive (CBI+) and CBI-negative (CBI-) groups. Marker performance was compared, and receiver operating characteristic analysis conducted. Pearson correlation assessed the agreement between BGA and laboratory measurements.

RESULTS

Among the 130 CSF samples, 11 were CBI+. Both cerebrospinal fluid lactate (cLac) and CLGR were significantly elevated in the CBI + group (P < 0.001). The area under the curve for cLac and CLGR was 0.990 and 0.994, respectively. Using a cutoff of 6.0 mmol/L, cLac demonstrated sensitivity of 100%, specificity of 93.3%, positive predictive value of 57.9%, negative predictive value of 100%, and diagnostic accuracy of 93.9%. CLGR ≥20.0 showed even higher accuracy: 100.0% sensitivity, 98.6% specificity, 84.6% positive predictive value, 100% negative predictive value, and overall accuracy of 98.5%. Both markers maintained excellent performance in blood-stained CSF. BGA measurements correlated well with laboratory results (r = 0.980 and 0.999, respectively, P < 0.001).

CONCLUSIONS

CLac levels ≥6.0 mmol/L and CLGR ≥20.0 accurately identified CBI in neurosurgical patients, with CLGR exhibiting superior efficacy. The potential for instant BGA measurement suggests promise for point-of-care testing.

摘要