Nykamp Moritz, Becker Fabian, Hoelzmann Philipp
Freie Universität Berlin, Institute of Geographical Sciences, Malteserstr. 74-100, 12249 Berlin, Germany.
MethodsX. 2024 Aug 30;13:102934. doi: 10.1016/j.mex.2024.102934. eCollection 2024 Dec.
The total organic carbon (TOC) concentration of particulate samples is a key parameter to characterize soils and sediments. To demonstrate the applicability and reliability of a modified sample preparation method for the direct measurement of TOC contents in suspended particulate samples, we analyzed five certified reference materials (CRMs) with varying TOC concentrations using a Shimadzu TOC-L CPH analyzer. Measured values were calibrated with a multi-point curve that cover the full range of the expected TOC concentrations and the results were validated using statistical values and measures. The method validation reveals that the measurements are accurate and precise for CRMs from marine and soil contexts, but show a low accuracy for the CRM containing polycyclic aromatic hydrocarbons (PAHs). This demonstrates the applicability and reliability of the modified preparation method for direct TOC determination of suspended particulate samples. Therefore, it is relevant for a broader community, beyond geosciences, and for users employing devices of other manufacturers to analyze TOC in suspended particulate samples.•Modified preparation method uses reduced sample weights and yields accurate and precise results.•Cost-efficient and environmentally friendly alternative: reduces waste by saving acid and ultrapure water.•Avoids incomplete dissolution of dolomite by heating acidified samples.
颗粒样品的总有机碳(TOC)浓度是表征土壤和沉积物的关键参数。为了证明一种改进的样品制备方法用于直接测量悬浮颗粒样品中TOC含量的适用性和可靠性,我们使用岛津TOC-L CPH分析仪分析了五种TOC浓度不同的有证标准物质(CRM)。测量值通过覆盖预期TOC浓度全范围的多点曲线进行校准,并使用统计值和测量方法对结果进行验证。方法验证表明,对于海洋和土壤环境的CRM,测量结果准确且精确,但对于含有多环芳烃(PAH)的CRM,准确性较低。这证明了改进的制备方法用于直接测定悬浮颗粒样品中TOC的适用性和可靠性。因此,它对于地球科学以外的更广泛群体以及使用其他制造商设备分析悬浮颗粒样品中TOC的用户具有相关性。
•改进的制备方法使用减少的样品重量并产生准确和精确的结果。
•经济高效且环保的替代方法:通过节省酸和超纯水减少浪费。
•通过加热酸化样品避免白云石不完全溶解。