使用模拟患者研究评估平面门控血池处理软件的准确性。

Evaluating the accuracy of planar gated blood pool processing software using simulated patient studies.

作者信息

Pieters H, van Staden J A, du Raan H, Nel M G, Engelbrecht G H J

机构信息

Nuclear Medicine Unit, Klerksdorp/Tshepong Hospital Complex, Klerksdorp, 2571, South Africa.

Department of Medical Physics, University of the Free State, Bloemfontein, 9301, South Africa.

出版信息

Heliyon. 2024 Sep 2;10(17):e37299. doi: 10.1016/j.heliyon.2024.e37299. eCollection 2024 Sep 15.

Abstract

Planar gated blood pool (GBP-P) radionuclide imaging is a valuable non-invasive technique for assessing left ventricular ejection fraction (LVEF). Serial cardiac imaging can be performed to monitor the potential decline in LVEF among patients undergoing cardiotoxic chemotherapy. Consequently, accurate LVEF determination becomes paramount. While commercial software programs have enhanced the LVEF values' reproducibility, concerns remain regarding their accuracy. This study aimed to generate a database of GBP-P studies with known LVEF values using Monte Carlo simulations and to assess LVEF values' accuracy using four commercial software programs. We utilised anthropomorphic 4D-XCAT models to generate 64 clinically realistic GBP-P studies with Monte Carlo simulations. Four commercial software programs (Alfanuclear, Siemens, General Electric Xeleris, and Mediso Tera-Tomo) were used to process these simulated studies. The accuracy and reproducibility of the LVEF values determined with these software programs and the intra- and inter-observer reproducibility of the LVEF values were assessed. Our study revealed a strong correlation between LVEF values calculated by the software programs and the true LVEF values derived from the 4D-XCAT models. However, all the software programs slightly underestimated LVEF at lower LVEF values. Intra- and inter-observer reliability for LVEF measurements was excellent. Accurate LVEF assessment is crucial for determining the patient's cardiac function before initiating and during chemotherapy treatment. The observed underestimation, particularly at lower LVEF values, emphasises the need for the accurate and reproducible determination of these values to avoid excluding suitable candidates for chemotherapy. The software programs' excellent intra- and inter-observer reliability highlights their potential to reduce subjectivity when using the semi-automatic processing option. This study confirms the accuracy and reliability of these commercial software programs in determining LVEF values from simulated GBP-P studies. Future research should investigate strategies to mitigate the underestimation biases and extend findings to diverse patient populations. Gated blood pool studies, left ventricular ejection fraction, Monte Carlo simulations, 4D-XCAT models.

摘要

平面门控血池(GBP-P)放射性核素成像术是评估左心室射血分数(LVEF)的一项重要无创技术。可进行系列心脏成像,以监测接受心脏毒性化疗患者的LVEF潜在下降情况。因此,准确测定LVEF至关重要。虽然商业软件程序提高了LVEF值的可重复性,但对其准确性仍存在担忧。本研究旨在使用蒙特卡洛模拟生成具有已知LVEF值的GBP-P研究数据库,并使用四个商业软件程序评估LVEF值的准确性。我们利用人体4D-XCAT模型通过蒙特卡洛模拟生成64项临床逼真的GBP-P研究。使用四个商业软件程序(Alfanuclear、西门子、通用电气Xeleris和Mediso Tera-Tomo)处理这些模拟研究。评估了用这些软件程序确定的LVEF值的准确性和可重复性以及LVEF值在观察者内和观察者间的可重复性。我们的研究揭示了软件程序计算的LVEF值与源自4D-XCAT模型的真实LVEF值之间存在很强的相关性。然而,所有软件程序在较低LVEF值时都略微低估了LVEF。LVEF测量的观察者内和观察者间可靠性极佳。准确评估LVEF对于在化疗治疗开始前和治疗期间确定患者的心功能至关重要。观察到的低估现象,尤其是在较低LVEF值时,强调了准确和可重复确定这些值的必要性,以避免排除合适的化疗候选者。软件程序出色的观察者内和观察者间可靠性突出了其在使用半自动处理选项时降低主观性的潜力。本研究证实了这些商业软件程序在从模拟GBP-P研究中确定LVEF值方面的准确性和可靠性。未来的研究应探讨减轻低估偏差的策略,并将研究结果扩展到不同患者群体。门控血池研究、左心室射血分数、蒙特卡洛模拟、4D-XCAT模型

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/1783/11408074/acad2d4aa592/gr1.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验