Kaiser Patricia, Kramer Udo, Rosenthal Hannah, Genz Christian, Weiss Nathalie, Schellenberg Ingo, Spannagl Michael
Instand e.V., Duesseldorf, Germany.
Hart Biologicals Ltd., Hartlepool, UK.
Clin Chem Lab Med. 2024 Sep 24;63(3):552-558. doi: 10.1515/cclm-2024-0822. Print 2025 Feb 25.
Until now, the external quality assessment (EQA) of glucose point-of-care testing (POCT) has lacked a high quality, suitable and commutable control material to assess measurement accuracy. Here we present a concept for determining the accuracy of glucose measurements, which uses human whole blood and does not require stabilising agents.
This new generation of quality control samples uses a bead that contains a specific amount of glucose. The bead is then dissolved in a whole blood matrix by the EQA participant immediately before the POCT. We analysed its suitability as an EQA material with respect to its reproducibility, homogeneity and stability, and applied it in an EQA pilot study. The glucose target value was determined using the reference measurement procedure and served as an evaluation criterion for the accuracy of the EQA survey results.
The homogeneity and stability of the new control material fulfilled the quality requirements of ISO 17043. Based on the reference measurement value for glucose, the results of the pilot EQA scheme showed a pass rate of 84.6 % for the participating POCT devices. The acceptance limit was a 15 % permitted deviation from the target value according to Rili-BAEK. All of the device collectives deviated from the target value by 0-4.4 % with the exception of one device type, which deviated by 21 %.
The new concept offers, for the first-time, whole blood-based trueness controls for glucose POCT analysis for external quality assurance. The concept does not require the addition of any stabilising reagent and is easy to use.
到目前为止,即时检测(POCT)血糖仪的外部质量评估(EQA)一直缺乏一种高质量、合适且可互换的对照物质来评估测量准确性。在此,我们提出一种用于确定葡萄糖测量准确性的概念,该概念使用人体全血且不需要稳定剂。
新一代质量控制样本使用含有特定量葡萄糖的珠子。然后,EQA参与者在进行POCT之前立即将珠子溶解在全血基质中。我们分析了其作为EQA材料在重复性、均匀性和稳定性方面的适用性,并将其应用于一项EQA试点研究。使用参考测量程序确定葡萄糖目标值,并将其作为EQA调查结果准确性的评估标准。
新对照物质的均匀性和稳定性满足ISO 17043的质量要求。基于葡萄糖的参考测量值,试点EQA方案的结果显示,参与的POCT设备的合格率为84.6%。根据Rili-BAEK,接受限为允许偏离目标值15%。除一种设备类型偏差21%外,所有设备组与目标值的偏差为0-4.4%。
这一新概念首次为葡萄糖POCT分析的外部质量保证提供了基于全血的准确性对照。该概念不需要添加任何稳定试剂,且易于使用。