Reiniger G, Rudolph W
Herz. 1985 Oct;10(5):305-11.
Subsequent to the finding of a rapidly attenuated anti-ischemic effect on use of high-dose nitroglycerin patch treatment in patients with coronary artery disease as well as contradictory results obtained with low-dose patch treatment, there are still no reliable guidelines for the use of transdermal nitroglycerin patch therapy. Accordingly, this study was carried out in ten patients with angiographically-documented coronary artery disease, stable exercise-induced angina pectoris and reproducible ST-segment depression, to assess the extent and duration of antianginal and anti-ischemic action of nitroglycerin patch treatment at an intermediate dosage of 15 mg/24 hours after the initial application, after renewed patch application on the second day and after patch application on the third day which had been preceded by a ten-hour patch-free interval in the night. At 2.5 hours after initial application of the nitroglycerin patch there was a 64% reduction in exercise-induced ST-segment depression from 2.6 mm +/- 0.21 (SEM) to 0.93 mm +/- 0.29 (p less than 0.001) (Figures 1 and 2). The exercise capacity to onset of 1 mm ST-segment depression was increased 132% from 191 watt X min +/- 15 (SEM) to 442 watt X min +/- 47 (p less than 0.001), (Figure 3). Eight hours after the initial patch application, exercise-induced ST-segment depression was reduced 37% from 2.6 mm +/- 0.19 to 1.65 mm +/- 0.18 (p less than 0.01); the exercise capacity to onset of 1 mm ST-segment depression was increased 39% from 178 watt X min +/- 18 to 245 watt X min +/- 19 (p less than 0.05).(ABSTRACT TRUNCATED AT 250 WORDS)
在发现高剂量硝酸甘油贴片治疗对冠心病患者的抗缺血作用迅速减弱,以及低剂量贴片治疗结果相互矛盾之后,目前仍没有关于经皮硝酸甘油贴片疗法使用的可靠指南。因此,本研究针对十名经血管造影证实患有冠心病、稳定的运动诱发型心绞痛且ST段压低可再现的患者进行,以评估在初始应用后、第二天重新贴片应用后以及第三天贴片应用(前一晚有10小时无贴片间隔)后,15毫克/24小时中等剂量硝酸甘油贴片治疗的抗心绞痛和抗缺血作用的程度及持续时间。在初始应用硝酸甘油贴片后2.5小时,运动诱发的ST段压低从2.6毫米±0.21(标准误)降至0.93毫米±0.29,降低了64%(p<0.001)(图1和图2)。运动至ST段压低1毫米开始的运动能力从191瓦·分钟±15(标准误)增加到442瓦·分钟±47,增加了132%(p<0.001)(图3)。初始贴片应用8小时后,运动诱发的ST段压低从2.6毫米±0.19降至1.65毫米±0.18,降低了37%(p<0.01);运动至ST段压低1毫米开始的运动能力从178瓦·分钟±18增加到245瓦·分钟±19,增加了39%(p<0.05)。(摘要截断于250字)