Friedburg H, Hennig J, Wenz W
Radiologe. 1985 Sep;25(9):426-9.
As an alternative to costly MRI systems with superconductive magnets, first experiences with an iron-shielded resistive system (Tomikon BMT 1100, Bruker Medizintechnik--Rheinstetten) are reported. The magnetic field strength is 0.235 T, corresponding to a proton resonance frequency of 10 MHz. Our system is used mainly for research, the image quality being comparable with that of superconducting systems. Especially favourable is the low magnetic stray field of the self-shielding system. It allows unproblematic handling of patients in a clinical environment. Economic aspects are not discussed.
作为具有超导磁体的昂贵MRI系统的替代方案,本文报道了首次使用铁屏蔽电阻系统(Tomikon BMT 1100,布鲁克医疗技术公司——莱茵施泰滕)的经验。磁场强度为0.235 T,对应质子共振频率为10 MHz。我们的系统主要用于研究,其图像质量与超导系统相当。特别有利的是自屏蔽系统的低杂散磁场。这使得在临床环境中对患者的处理毫无问题。本文未讨论经济方面的问题。