Nuccitelli Richard, Litt Jeffrey, Moss Kevin, Hinman Dale, Getsin Ilya, Uecker Darrin, Stoffregen Bill, Knape William A
Pulse Biosciences, Inc., Hayward, California, USA.
Northstar Preclinical and Pathology Services, Lake Elmo, Minnesota, USA.
Bioelectricity. 2024 Sep 16;6(3):174-180. doi: 10.1089/bioe.2024.0015. eCollection 2024 Sep.
The purpose of this study was to demonstrate the safety and performance of the CellFX Percutaneous Electrode for delivering nanosecond pulsed field ablation (nsPFA) energy to soft tissues. Three different porcine tissue types were treated, namely, liver, kidney, and skeletal muscle, at treatment levels of three times greater than clinical treatment levels. The histological characteristics of the ablation zone for each of these tissues compared with that of radiofrequency (RF) ablation on day 0 and 2 days post-treatment. Ablation zone dimensions were measured during gross necropsy after tetrazolium chloride staining and compared between the nsPFA and RF groups at 2 days post-ablation. The CellFX system successfully achieved ablation and necrosis of all treatment sites in all target tissues. No evidence of thermal effects or collagen degeneration was found at any of the nsPFA treatment sites. Overall systemic tolerability was evidenced by the absence of clinically significant changes in urinalysis and serum chemistry before and after treatments.
本研究的目的是证明CellFX经皮电极在向软组织输送纳秒脉冲电场消融(nsPFA)能量方面的安全性和性能。对三种不同的猪组织类型进行了治疗,即肝脏、肾脏和骨骼肌,治疗水平是临床治疗水平的三倍。将这些组织中每个组织的消融区组织学特征与治疗后第0天和第2天的射频(RF)消融特征进行比较。在氯化四氮唑染色后的大体尸检期间测量消融区尺寸,并在消融后第2天比较nsPFA组和RF组之间的消融区尺寸。CellFX系统成功实现了所有靶组织中所有治疗部位的消融和坏死。在任何nsPFA治疗部位均未发现热效应或胶原变性的证据。治疗前后尿液分析和血清化学指标无临床显著变化,证明了总体全身耐受性良好。