Steiner Joseph R, Morrison Courtney K, Vaya Mayur, Bevins Nicholas, Christophel Jeremy, Vanderhoek Matt
Department of Radiology, University of Chicago, Chicago, Illinois, USA.
Department of Diagnostic Radiology and Nuclear Medicine, RUSH University Medical Center, Chicago, Illinois, USA.
J Appl Clin Med Phys. 2024 Dec;25(12):e14536. doi: 10.1002/acm2.14536. Epub 2024 Oct 9.
Fluoroscopy uses collimators to limit the radiation field size. Collimators are often evaluated annually during equipment performance evaluations to maintain compliance with regulatory and/or accreditation bodies. A method to evaluate and quantify fluoroscopy collimator performance was developed.
A radiation field and displayed image measurement device consisting of radiopaque rulers and radiochromic film strips was placed on the x-ray source assembly exit window to evaluate fluoroscopy collimator performance. This method was used to evaluate collimator performance on 79 fluoroscopic imaging systems including fixed C-arms, mobile C-arms, mini C-arms, and radiographic fluoroscopic systems.
The excess length (EL), excess width (EW), and sum EL + EW of the radiation field relative to the displayed image were measured and compared to the limits specified in 21CFR1020.32. Four systems exceeded these limits. Placing the radiation measurement device at the x-ray source assembly exit window relative to the image receptor cover increased the film exposure rate by a factor up to 14.6. The time required to set up and complete the fluoroscopy collimator performance measurements using this method ranged from 5 to 10 min.
This method provides an easily implemented quantitative measure of fluoroscopy system collimator performance that satisfies regulatory and accreditation body requirements.
荧光透视检查使用准直器来限制辐射野大小。在设备性能评估期间,通常每年对准直器进行评估,以保持符合监管和/或认证机构的要求。已开发出一种评估和量化荧光透视准直器性能的方法。
将由不透射线的尺子和放射变色胶片条组成的辐射野和显示图像测量装置放置在X射线源组件的出射窗口上,以评估荧光透视准直器的性能。该方法用于评估79台荧光透视成像系统的准直器性能,这些系统包括固定C形臂、移动C形臂、迷你C形臂和放射荧光透视系统。
测量了辐射野相对于显示图像的超长(EL)、超宽(EW)以及EL+EW之和,并与21 CFR1020.32中规定的限值进行比较。有4个系统超出了这些限值。将辐射测量装置相对于图像接收器罩放置在X射线源组件的出射窗口处,使胶片曝光率提高了14.6倍。使用该方法设置并完成荧光透视准直器性能测量所需的时间为5至10分钟。
该方法提供了一种易于实施的荧光透视系统准直器性能定量测量方法,满足监管和认证机构的要求。