Impact of Skin Pigmentation on Pulse Oximetry Blood Oxygenation and Wearable Pulse Rate Accuracy: Systematic Review and Meta-Analysis.

机构信息

University of Michigan Medical School, Ann Arbor, MI, United States.

Verily Life Sciences LLC, South San Francisco, CA, United States.

出版信息

J Med Internet Res. 2024 Oct 10;26:e62769. doi: 10.2196/62769.

Abstract

BACKGROUND

Photoplethysmography (PPG) is a technology routinely used in clinical practice to assess blood oxygenation (SpO) and pulse rate (PR). Skin pigmentation may influence accuracy, leading to health outcomes disparities.

OBJECTIVE

This systematic review and meta-analysis primarily aimed to evaluate the accuracy of PPG-derived SpO and PR by skin pigmentation. Secondarily, we aimed to evaluate statistical biases and the clinical relevance of PPG-derived SpO and PR according to skin pigmentation.

METHODS

We identified 23 pulse oximetry studies (n=59,684; 197,353 paired SpO-arterial blood observations) and 4 wearable PR studies (n=176; 140,771 paired PPG-electrocardiography observations). We evaluated accuracy according to skin pigmentation group by comparing SpO accuracy root-mean-square values to the regulatory threshold of 3% and PR 95% limits of agreement values to +5 or -5 beats per minute (bpm), according to the standards of the American National Standards Institute, Association for the Advancement of Medical Instrumentation, and the International Electrotechnical Commission. We evaluated biases and clinical relevance using mean bias and 95% CI.

RESULTS

For SpO, accuracy root-mean-square values were 3.96%, 4.71%, and 4.15%, and pooled mean biases were 0.70% (95% CI 0.17%-1.22%), 0.27% (95% CI -0.64% to 1.19%), and 1.27% (95% CI 0.58%-1.95%) for light, medium, and dark pigmentation, respectively. For PR, 95% limits of agreement values were from -16.02 to 13.54, from -18.62 to 16.84, and from -33.69 to 32.54, and pooled mean biases were -1.24 (95% CI -5.31 to 2.83) bpm, -0.89 (95% CI -3.70 to 1.93) bpm, and -0.57 (95% CI -9.44 to 8.29) bpm for light, medium, and dark pigmentation, respectively.

CONCLUSIONS

SpO and PR measurements may be inaccurate across all skin pigmentation groups, breaching U.S. Food and Drug Administration guidance and industry standard thresholds. Pulse oximeters significantly overestimate SpO for both light and dark skin pigmentation, but this overestimation may not be clinically relevant. PRs obtained from wearables exhibit no statistically or clinically significant bias based on skin pigmentation.

摘要

背景

光体积描记法(PPG)是一种在临床实践中常规用于评估血氧饱和度(SpO)和脉搏率(PR)的技术。皮肤色素沉着可能会影响准确性,导致健康结果存在差异。

目的

本系统评价和荟萃分析主要旨在评估根据皮肤色素沉着评估 PPG 衍生的 SpO 和 PR 的准确性。其次,我们旨在根据皮肤色素沉着评估 PPG 衍生的 SpO 和 PR 的统计偏差和临床相关性。

方法

我们确定了 23 项脉搏血氧仪研究(n=59684;197353 对 SpO-动脉血观察)和 4 项可穿戴 PR 研究(n=176;140771 对 PPG-心电图观察)。我们根据美国国家标准协会、医疗仪器促进协会和国际电工委员会的标准,通过比较 SpO 准确性均方根值与 3%的监管阈值以及 PR 95%一致性界限值与+5 或-5 次/分钟(bpm),根据皮肤色素沉着组评估准确性。我们使用平均偏差和 95%置信区间评估偏差和临床相关性。

结果

对于 SpO,准确性均方根值分别为 3.96%、4.71%和 4.15%,平均偏差分别为 0.70%(95%置信区间 0.17%-1.22%)、0.27%(95%置信区间-0.64%-1.19%)和 1.27%(95%置信区间 0.58%-1.95%),用于轻度、中度和重度色素沉着。对于 PR,95%一致性界限值分别为-16.02 至 13.54、-18.62 至 16.84 和-33.69 至 32.54,平均偏差分别为-1.24(95%置信区间-5.31 至 2.83)bpm、-0.89(95%置信区间-3.70 至 1.93)bpm 和-0.57(95%置信区间-9.44 至 8.29)bpm,用于轻度、中度和重度色素沉着。

结论

在所有皮肤色素沉着组中,SpO 和 PR 测量值可能不准确,不符合美国食品和药物管理局的指导意见和行业标准阈值。脉搏血氧仪对浅色和深色皮肤色素沉着的 SpO 均存在显著高估,但这种高估可能不具有临床意义。可穿戴设备获得的 PR 没有基于皮肤色素沉着的统计学或临床显著偏差。