Latta Peter, Jiřík Radovan, Vitouš Jiří, Macíček Ondřej, Vojtíšek Lubomír, Rektor Ivan, Standara Michal, Křístek Jan, Starčuk Zenon
Central European Institute of Technology, Masaryk University, Brno, Czech Republic.
Institute of Scientific Instruments of the Czech Academy of Sciences, Brno, Czech Republic.
Magn Reson Med. 2025 Mar;93(3):1285-1297. doi: 10.1002/mrm.30355. Epub 2024 Oct 22.
The aim of this study was to develop a simple, robust, and easy-to-use calibration procedure for correcting misalignments in rosette MRI k-space sampling, with the objective of producing images with minimal artifacts.
Quick automatic calibration scans were proposed for the beginning of the measurement to collect information on the time course of the rosette acquisition trajectory. A two-parameter model was devised to match the measured time-varying readout gradient delays and approximate the actual rosette sampling trajectory. The proposed calibration approach was implemented, and performance assessment was conducted on both phantoms and human subjects.
The fidelity of phantom and in vivo images exhibited significant improvement compared with uncorrected rosette data. The two-parameter calibration approach also demonstrated enhanced precision and reliability, as evidenced by quantitative relaxometry analyses.
Adequate correction of data sampling is a crucial step in rosette MRI. The presented experimental results underscore the robustness, ease of implementation, and suitability for routine experimental use of the proposed two-parameter rosette trajectory calibration approach.
本研究的目的是开发一种简单、稳健且易于使用的校准程序,用于校正玫瑰花结MRI k空间采样中的错位,目标是生成伪影最少的图像。
建议在测量开始时进行快速自动校准扫描,以收集有关玫瑰花结采集轨迹时间进程的信息。设计了一个双参数模型来匹配测量的随时间变化的读出梯度延迟,并近似实际的玫瑰花结采样轨迹。实施了所提出的校准方法,并在体模和人体受试者上进行了性能评估。
与未校正的玫瑰花结数据相比,体模和体内图像的保真度有显著提高。双参数校准方法还显示出更高的精度和可靠性,定量弛豫分析证明了这一点。
数据采样的充分校正是玫瑰花结MRI中的关键步骤。所呈现的实验结果强调了所提出的双参数玫瑰花结轨迹校准方法的稳健性、易于实施性以及适用于常规实验使用的特性。