Mader Julia K, Baumstark Annette, Tüting Johannes, Sokol Günter, Schuebel Ruth, Tong Yuhong, Roetschke Julia, Slingerland Robbert J
Division of Endocrinology and Diabetology, Medical University of Graz, Graz, Austria.
Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft GmbH an Der Universität Ulm, Ulm, Germany.
Diabetes Ther. 2024 Dec;15(12):2525-2535. doi: 10.1007/s13300-024-01665-9. Epub 2024 Oct 28.
A sizeable minority of commercially available blood glucose monitoring (BGM) systems fail to satisfy regulatory accuracy requirements, such as ISO 15197:2013, after approval. This study assessed whether the BGMs tested could consistently meet these ISO requirements by investigating their accuracy in a non-standardized setting.
In this 18-month post-market performance study, using the ISO criteria, healthcare professionals tested the accuracy of four CE-marked BGM systems (Roche Diabetes Care, Mannheim, Germany) on European adults with diabetes mellitus. ISO criteria included 95% of blood glucose (BG) values being within ± 15 mg/dl of a reference measurement for BG < 100 mg/dl or ± 15% for BG ≥ 100 mg/dl and, in the Parkes Consensus Error grid for type 1 diabetes comparing capillary BGM measurements versus reference method, 99% of BG values falling within zone A (no effect on clinical action or outcome) and zone B (altered clinical action with little or no effect on clinical outcome).
BGM readings were obtained from 1650 participants, and the number of readings per BGM system was between 1712 and 2376. The percentage of BGM readings that fell within ISO 15197:2013 limits ranged from 99.4 to 99.9%. For all meter types, 100% of data points fell within zone A or zone B, and most data points for each meter (≥ 99.9%) were in zone A.
All four CE-marked BGM models showed results within the accuracy limits defined by ISO 15197 in a non-standardized setting and thus consistently met regulatory accuracy requirements.
相当一部分市售血糖监测(BGM)系统在获批后未能满足监管准确性要求,如ISO 15197:2013。本研究通过在非标准化环境中调查其准确性,评估所测试的BGM能否始终满足这些ISO要求。
在这项为期18个月的上市后性能研究中,医疗专业人员使用ISO标准,对四种CE标志的BGM系统(德国曼海姆罗氏糖尿病护理公司)在欧洲成年糖尿病患者中进行了准确性测试。ISO标准包括95%的血糖(BG)值在BG<100mg/dl时相对于参考测量值±15mg/dl范围内,或BG≥100mg/dl时±15%范围内,并且在1型糖尿病的帕克斯共识误差网格中,将毛细血管BGM测量值与参考方法进行比较时,99%的BG值落在A区(对临床行动或结果无影响)和B区(改变临床行动但对临床结果影响很小或无影响)内。
从1650名参与者中获取了BGM读数,每个BGM系统的读数数量在1712至2376之间。落在ISO 15197:2013限值范围内的BGM读数百分比在99.4%至99.9%之间。对于所有血糖仪类型,100%的数据点落在A区或B区内,并且每个血糖仪的大多数数据点(≥99.9%)在A区内。
所有四种CE标志的BGM型号在非标准化环境中均显示出在ISO 15197定义的准确性限值内的结果,因此始终满足监管准确性要求。