The development of an implantable electrochemical glucose sensor: response to glucose in bovine serum ultrafiltrate.

作者信息

Guyton J R, Soeldner J S, Giner J, Colton C K, Morris R S

出版信息

Horm Metab Res Suppl. 1979(8):24-6.

PMID:395084
Abstract

A glucose sensor based on organic oxidation reactions at a platinum electrode is being developed as the key component for an implantable artificial beta cell for diabetic patients. Sensitivity of a membrane-covered platinum electrode to changing glucose concentration in bovine serum ultrafiltrate has been demonstrated, with current response as high as 60 microamperes per 100 mg/dl change in glucose concentration. Reproducibility of measurements probably will depend on satisfactory mechanical assembly of the membrane-electrode combination. An approach to selectivity based on multiple current measurements is suggested.

摘要

一种基于铂电极上有机氧化反应的葡萄糖传感器正在被开发,作为糖尿病患者可植入人工β细胞的关键组件。已证明膜覆盖铂电极对牛血清超滤液中葡萄糖浓度变化的敏感性,葡萄糖浓度每变化100mg/dl,电流响应高达60微安。测量的可重复性可能取决于膜电极组合的令人满意的机械组装。提出了一种基于多次电流测量的选择性方法。