Beck J W, Tatum J L, Cobb F R, Harris C C, Goodrich J K
J Nucl Med. 1979 Dec;20(12):1294-300.
A method is presented for calculating a background image to be subtracted from TI-201 myocardial perfusion images. The method was derived from experimental measurements of background components in which hearts of animals injected with TI-201 were replaced with hearts from nonradioactive animals. The algorithm generates a background image that accounts for TI-201 activity in surrounding tissues and within the cardiac chamber. Comparison of the computer-generated background images with background images of the experimental models showed a mean difference of about 3% (range 1-6%). Clinical images using this method of background generation and subtraction are presented.