Centro Universitário Christus - Fortaleza (CE) - Brazil.
Acta Cir Bras. 2024 Nov 8;39:e397624. doi: 10.1590/acb397624. eCollection 2024.
To develop and validate a trans oral endoscopic thyroidectomy vestibular approach (TOETVA) simulator.
The first phase of the project consisted of designing and developing a transoral thyroid surgery simulator based on real surgeries. The product has the oral cavity for attaching the three trocars and the cervical part containing the thyroid and adjacent structures. In the second phase, the simulator was validated by specialists who performed an endoscopic thyroidectomy procedure. They all filled a questionnaire about the simulator and the simulation based on the Likert scale.
The simulator consists of a console similar to a human bust and a high-resolution camera system connected to a 22-inch monitor. The simulator had excellent results in the visual evaluation (face validity), with 100% of responses between good and excellent for the following characteristics: synthetic structures, design, visibility of the surgery field, resistance, resilience, fulcrum effect, ergonomics, surgical material, and practicality. The last three were rated higher, with more votes for excellent. For content validity, the items that received the best ratings were, precisely, the steps relating to the surgical procedure: opening the intermuscular midline, isthmotomy, and thyroidectomy.
The thyroidectomy training box showed great ability to simulate a TOETVA, with satisfactory evaluations concerning its visual and content validation.
开发和验证经口内镜甲状腺手术 Vestibular 入路(TOETVA)模拟器。
该项目的第一阶段包括设计和开发基于真实手术的经口甲状腺手术模拟器。该产品具有用于连接三个 trocar 的口腔和包含甲状腺和相邻结构的颈部部分。在第二阶段,专家们通过内窥镜甲状腺切除术程序对模拟器进行了验证。他们根据李克特量表填写了一份关于模拟器和模拟的问卷。
模拟器由类似于人体半身像的控制台和连接到 22 英寸监视器的高分辨率摄像系统组成。模拟器在视觉评估(表面有效性)方面取得了出色的结果,以下特征的反应均为良好到优秀之间:合成结构、设计、手术视野的可视性、阻力、弹性、枢轴效应、人体工程学、手术材料和实用性。后三个特征的评分更高,有更多的优秀评价。在内容有效性方面,评分最高的项目恰恰是与手术过程相关的步骤:打开肌肉间中线、峡部切开术和甲状腺切除术。
甲状腺切除术训练箱具有出色的模拟经口内镜甲状腺手术 Vestibular 入路的能力,在视觉和内容验证方面得到了令人满意的评价。