Li Xiuquan, Deng Jia, Wu Xiangyang, Yang Hang, Huang Dengdian
Department of Radiation Oncology, Shaanxi Provincial Tumor Hospital, Xi'an, Shaanxi, China.
Department of Oncology, The First Affiliated Hospital of Chongqing Medical University, Chongqing, China.
Radiat Environ Biophys. 2025 Mar;64(1):55-65. doi: 10.1007/s00411-024-01097-w. Epub 2024 Nov 14.
This study aimed to comprehensively investigate the influence of multi-leaf collimator (MLC) position errors on both the clinical absolute dose distribution and Gamma passing rate (%GP) in intensity-modulated radiation therapy (IMRT) plans for breast cancer. Additionally, the correlation between %GP and the clinical absolute dose relative difference (%DE) caused by MLC position errors was analysed. Ten IMRT plans for breast cancer were randomly selected. Systematic and random MLC position errors were introduced into DICOM files representing the investigated treatment plans by modifying the plan files and adjusting the MLC positions. Systematic errors were categorized as MLC opening errors, closing errors, and shift errors. The %DE in the tumour planning target volume (PTV) and organs at risk (OARs) caused by MLC errors were statistically analyzed using dose-volume histogram (DVH) analysis. The ArcCHECK quality assurance (QA) system was used to detect the %GP differences between baseline plans and plans with MLC errors. The correlation between %GP and %DE was obtained using linear regression methods. The results of this study indicate that MLC opening and closing errors have a significant impact on %DE and %GP in IMRT plans for breast cancer. Opening and closing errors can be detected at a gamma level of 3%/2 mm, if error values are greater than or equal to 0.5 mm, and %GP can predict DVH dosimetric changes caused by MLC opening and closing errors. It is concluded that DVH-based verification of IMRT plans can serve as an adjunct method to Gamma analysis to improve QA accuracy for breast cancer cases. Additionally, it is concluded that greater attention should be given to MLC leaf opening and closing errors in clinical practice.
本研究旨在全面调查多叶准直器(MLC)位置误差对乳腺癌调强放射治疗(IMRT)计划中临床绝对剂量分布和伽马通过率(%GP)的影响。此外,还分析了%GP与由MLC位置误差引起的临床绝对剂量相对差异(%DE)之间的相关性。随机选择了10个乳腺癌IMRT计划。通过修改计划文件和调整MLC位置,将系统和随机的MLC位置误差引入代表所研究治疗计划的DICOM文件中。系统误差分为MLC打开误差、关闭误差和移位误差。使用剂量体积直方图(DVH)分析对由MLC误差引起的肿瘤计划靶体积(PTV)和危及器官(OARs)中的%DE进行统计分析。使用ArcCHECK质量保证(QA)系统检测基线计划与存在MLC误差的计划之间的%GP差异。使用线性回归方法获得%GP与%DE之间的相关性。本研究结果表明,MLC打开和关闭误差对乳腺癌IMRT计划中的%DE和%GP有显著影响。如果误差值大于或等于0.5 mm,在3%/2 mm的伽马水平下可以检测到打开和关闭误差,并且%GP可以预测由MLC打开和关闭误差引起的DVH剂量学变化。得出的结论是,基于DVH的IMRT计划验证可作为伽马分析的辅助方法,以提高乳腺癌病例的QA准确性。此外,得出的结论是,在临床实践中应更加关注MLC叶片的打开和关闭误差。