皮肤组织介电常数:一天中的时间和皮肤深度依赖性。

Skin tissue dielectric constant: Time of day and skin depth dependence.

作者信息

Mayrovitz Harvey N

机构信息

Department of Medical Education, Dr. Kiran C. Patel College of Allopathic Medicine, Nova Southeastern University, Fort Lauderdale, Florida, USA.

出版信息

Clin Physiol Funct Imaging. 2025 Jan;45(1):e12916. doi: 10.1111/cpf.12916. Epub 2024 Nov 17.

Abstract

BACKGROUND

Skin water measurements are used to investigate skin physiology, clinically study dermatological issues, and for conditions like diabetes, oedema, and lymphedema with measurements done at various times of day (TOD). One method used is skin's tissue dielectric constant (TDC), often clinically measured to a single depth of 2.5 mm. This report characterizes intraday variations measured to multiple depths to guide expected TOD and depth dependence.

MATERIALS AND METHODS

Twelve medical students self-measured TDC on their forearm to depths of 0.5, 1.5, 2.5, and 5.0 mm every 2 h from 08:00 to 24:00 h on 2 consecutive days. All were trained in the procedure.

RESULTS

TDC declined slightly from morning through evening, mostly at 0.5 mm for which TDC was reduced by 4%. TDC values were not related to participants' whole-body fat or water percentages. The TDC decrease was less at 1.5 mm where the reduction was 2.7%. At depths of 2.5 or 5.0 mm, there was no significant decrease in TOD.

CONCLUSION

Skin TDC shows a minor decreasing trend with an effect greater for shallower depths. In part, the clinical relevance of the findings relates to the confidence level associated with skin water estimates, based on TDC measurements, when measured at different TOD and depths during normal clinic hours. Based on the present data the TOD change is at most 4% and insignificant for measurement depths of 2.5 mm.

摘要

背景

皮肤水分测量用于研究皮肤生理学、临床研究皮肤病问题,以及用于糖尿病、水肿和淋巴水肿等病症,且在一天中的不同时间进行测量。常用的一种方法是测量皮肤的组织介电常数(TDC),临床上通常测量单一深度2.5毫米处的TDC。本报告描述了测量多个深度时的日内变化,以指导预期的一天中的时间(TOD)和深度依赖性。

材料与方法

12名医科学生在连续两天的08:00至24:00期间,每隔2小时在自己的前臂上自我测量TDC,测量深度为0.5、1.5、2.5和5.0毫米。所有人都接受了该操作的培训。

结果

TDC从早晨到晚上略有下降,主要是在0.5毫米深度处,TDC下降了4%。TDC值与参与者的全身脂肪或水分百分比无关。在1.5毫米深度处TDC下降较少,下降了2.7%。在2.5或5.0毫米深度处,一天中的时间(TOD)没有显著下降。

结论

皮肤TDC呈轻微下降趋势,对较浅深度的影响更大。部分而言,这些发现的临床相关性涉及基于TDC测量在正常门诊时间的不同一天中的时间(TOD)和深度进行测量时与皮肤水分估计相关的置信水平。根据目前的数据,一天中的时间(TOD)变化最多为4%,对于2.5毫米的测量深度而言不显著。