Weightman B, Light D
Biomaterials. 1986 Jan;7(1):20-4. doi: 10.1016/0142-9612(86)90083-9.
The effect of the surface finish of medical grade alumina and stainless steel on the wear rate of ultra-high molecular weight (UHMW) polyethylene was studied in a six station pin-on-plate wear screening device using bovine serum as lubricant. The wear rate decreased as the surface finish of alumina improved from 0.10 to 0.015 micron Ra, with no evidence of an increasing wear rate below a particular value of surface finish. Tests with surgical grade stainless steel showed a similar polyethylene wear rate around the 0.025 micron Ra level, but there was evidence of a higher wear rate with stainless steel than with alumina between approx. 0.035 and 0.075 micron Ra.
在一个六工位销盘式磨损筛选装置中,以牛血清作为润滑剂,研究了医用级氧化铝和不锈钢的表面光洁度对超高分子量(UHMW)聚乙烯磨损率的影响。随着氧化铝表面光洁度从0.10微米Ra提高到0.015微米Ra,磨损率降低,且没有证据表明在特定表面光洁度值以下磨损率会增加。外科级不锈钢的测试表明,在0.025微米Ra水平左右,聚乙烯的磨损率相似,但有证据表明,在约0.035至0.075微米Ra之间,不锈钢的磨损率高于氧化铝。