Suppr超能文献

长时间皮质内电刺激的组织病理学评估

Histopathologic evaluation of prolonged intracortical electrical stimulation.

作者信息

Agnew W F, Yuen T G, McCreery D B, Bullara L A

出版信息

Exp Neurol. 1986 Apr;92(1):162-85. doi: 10.1016/0014-4886(86)90132-9.

Abstract

Chronic stimulating microelectrodes fabricated from platinum-30% iridium (Pt-30%Ir) or activated iridium were implanted in assemblies of three in the left sensorimotor cortex of the cat and pulsed continuously at currents of 10 to 320 microA (100 to 3200 microC/cm2 X ph, 2 to 64 nC/ph) for periods of 24 h or for 23 h/day for 7 days. The microelectrodes had beveled tips with uninsulated geometric surface areas of 20 X 10(-6) cm2. Neuronal activity evoked by the focal stimulation was monitored by recording compound action potentials from the ipsilateral pyramidal tract. By this criterion neuronal activation thresholds were 5 to 15 microA (50 to 150 microC/cm2 X ph, 1 to 3 nC/ph) for both types of electrodes. Histologic evaluations of tissue surrounding the electrode tips were carried out by either light or electron microscopy. No neural damage was induced by 24 or 161 h of pulsing using either type of electrode at currents of 10 to 80 microA. Neural damage attributable to electrical stimulation per se was observed in a few sites pulsed with 320 microA (3200 microC/cm2 X ph, 64 nC/ph, 16 A/cm2) with Pt-30%Ir but not activated iridium electrodes of the same size. Electrode dissolution appears to be best correlated with charge density and current density. Dissolution of the Pt-30%Ir microelectrode tip was observed by scanning electron microscopy at charge densities as low as 200 microC/cm2 X ph (1 A/cm2), whereas erosion of activated iridium microelectrodes occurred only at the highest charge and current densities (3200 microC/cm2 X ph, 16 A/cm2). Thus, the activated iridium electrode is superior to Pt-30%Ir for chronic stimulations, from the standpoint of electrode tip stability, because with the former, in contrast to the alloy, detectable erosion occurred only at an intensity well above that required for activation of nearby neurons.

摘要

由铂 - 30%铱(Pt - 30%Ir)或活性铱制成的慢性刺激微电极以三个为一组植入猫的左侧感觉运动皮层,以10至320微安的电流(100至3200微库仑/平方厘米×脉冲,2至64纳库仑/脉冲)连续脉冲24小时,或每天脉冲23小时,持续7天。微电极的尖端为斜面,未绝缘的几何表面积为20×10⁻⁶平方厘米。通过记录同侧锥体束的复合动作电位来监测局灶性刺激诱发的神经元活动。根据这一标准,两种类型电极的神经元激活阈值均为5至15微安(50至150微库仑/平方厘米×脉冲,1至3纳库仑/脉冲)。通过光学或电子显微镜对电极尖端周围的组织进行组织学评估。使用任何一种电极在10至80微安的电流下脉冲24或161小时均未引起神经损伤。在用320微安(3200微库仑/平方厘米×脉冲,64纳库仑/脉冲,16安/平方厘米)脉冲的少数部位观察到了由电刺激本身引起的神经损伤,这些部位使用的是Pt - 30%Ir电极,而同样大小的活性铱电极则未出现这种情况。电极溶解似乎与电荷密度和电流密度最相关。通过扫描电子显微镜观察到,Pt - 30%Ir微电极尖端在低至200微库仑/平方厘米×脉冲(1安/平方厘米)的电荷密度下就出现溶解,而活性铱微电极仅在最高电荷和电流密度(3200微库仑/平方厘米×脉冲,16安/平方厘米)下才会出现腐蚀。因此,从电极尖端稳定性的角度来看,活性铱电极在慢性刺激方面优于Pt - 30%Ir电极,因为与合金电极相比,前者只有在远高于激活附近神经元所需强度时才会出现可检测到的腐蚀。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验