Suppr超能文献

临床实践中仅使用脑部磁共振成像的工作流程:质量保证和患者定位发生器之间的比较。

Brain MR-only workflow in clinical practice: A comparison among generators for quality assurance and patient positioning.

作者信息

Levardon Mathilde, Autret Damien, Le Dorze Thomas, Guillerminet Camille, Dufreneix Stéphane

机构信息

Department of Nuclear Medicine, Centre Hospitalier Universitaire d'Angers, Angers, France.

Department of Medical Physics, Institut de Cancérologie de l'Ouest, Angers, France.

出版信息

J Appl Clin Med Phys. 2025 Feb;26(2):e14583. doi: 10.1002/acm2.14583. Epub 2024 Nov 25.

Abstract

BACKGROUND AND PURPOSE

Routine quality control procedures are still required for sCT based on artificial intelligence (AI) to verify the performance of the generators. The aim of this study was to evaluate three generators based on AI or bulk density (BD) assignment for the patient-specific quality assurance (PSQA) of another AI-based generator in clinical routine. A patient positioning study based on 2D/2D kV-image comparing the performances of four sCT generators was also performed.

MATERIALS AND METHODS

On the four generators available commercially at our institution, one was chosen as the clinical one, and the three others were used for PSQA. Several dose metrics were calculated like the mean error, dose-volume histogram metrics, and 1%/1 mm gamma analysis. A comparison against CT was considered as a reference. Translations and rotations found during patient positioning based on sCT were compared to those based on CT.

RESULTS

Some of the metrics calculated against CT revealed patients outside the tolerances chosen (1% for point metrics; 90% for gamma pass rate). None of the generators was able to identify these outliers for all metrics studied. Performing a PSQA with other sCT generators introduced several false positives and false negatives. None of the generators was able to clearly identify, for all metrics studied, a true sCT failure caused by a metal implant. The smallest positioning deviations were found for the BD assignment sCT, the largest for the only AI generator not based on a T1 Dixon MR sequence.

CONCLUSIONS

PSQA of a sCT generator with another sCT generator should be performed with great care. Patient positioning is an important aspect to consider when evaluating a sCT generator. The results of this study should help medical physicists willing to set up a MR-only workflow for the brain based on a 2D/2D kV-image patient positioning.

摘要

背景与目的

基于人工智能(AI)的合成CT(sCT)仍需常规质量控制程序来验证其生成器的性能。本研究的目的是评估基于AI或体密度(BD)分配的三种生成器,用于临床常规中另一种基于AI的生成器的患者特定质量保证(PSQA)。还进行了一项基于二维/二维千伏图像比较四种sCT生成器性能的患者定位研究。

材料与方法

在我们机构可商购的四种生成器中,选择一种作为临床使用的生成器,另外三种用于PSQA。计算了几个剂量指标,如平均误差、剂量体积直方图指标和1%/1毫米伽马分析。与CT的比较被视为参考。将基于sCT的患者定位过程中发现的平移和旋转与基于CT的进行比较。

结果

一些针对CT计算的指标显示,患者超出了所选的公差范围(点指标为1%;伽马通过率为90%)。对于所研究的所有指标,没有一个生成器能够识别出这些异常值。使用其他sCT生成器进行PSQA会引入一些假阳性和假阴性。对于所研究的所有指标,没有一个生成器能够明确识别出由金属植入物导致的真正sCT故障。BD分配的sCT定位偏差最小,唯一不基于T1 Dixon MR序列的AI生成器定位偏差最大。

结论

使用另一种sCT生成器对sCT生成器进行PSQA时应格外谨慎。评估sCT生成器时,患者定位是一个重要的考虑因素。本研究结果应有助于医学物理学家基于二维/二维千伏图像患者定位为脑部建立仅使用磁共振成像的工作流程。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/427d/11799901/ab361666f968/ACM2-26-e14583-g005.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验