穿戴式惯性测量单元(IMU)测量健康老年人姿势摆动的结构效度及视觉剥夺的影响。

Construct Validity of a Wearable Inertial Measurement Unit (IMU) in Measuring Postural Sway and the Effect of Visual Deprivation in Healthy Older Adults.

机构信息

Department of Neurosciences, Biomedicine and Movement Sciences, University of Verona, 37131 Verona, Italy.

Department of Biomolecular Sciences, University of Urbino, 61029 Urbino, Italy.

出版信息

Biosensors (Basel). 2024 Nov 1;14(11):529. doi: 10.3390/bios14110529.

Abstract

Inertial Motor sensors (IMUs) are valid instruments for measuring postural sway but their ability to detect changes derived from visual deprivation in healthy older adults requires further investigations. We examined the validity and relationship of IMU sensor-derived postural sway measures compared to force plates for different eye conditions in healthy older adults (32 females, 33 males). We compared the relationship of the center of mass and center of pressure (CoM and CoP)-derived total length, root means square (RMS) distance, mean velocity, and 95% confidence interval ellipse area (95% CI ellipse area). In addition, we examined the relationship of the IMU sensor in discriminating between open- (EO) and closed-eye (EC) conditions compared to the force plate. A significant effect of the instruments and eye conditions was found for almost all the variables. Overall, EO and EC variables within (force plate r, from 0.38 to 0.78; IMU sensor r, from 0.36 to 0.69) as well as between (r from 0.50 to 0.88) instruments were moderately to strongly correlated. The EC:EO ratios of RMS distance and 95% CI ellipse area were not different between instruments, while there were significant differences between total length ( = 0.973) and mean velocity ( = 0.703). The ratios' correlation coefficients between instruments ranged from moderate (r = 0.65) to strong (r = 0.87). The IMU sensor offers an affordable, valid alternative to a force plate for objective, postural sway assessment.

摘要

惯性电机传感器 (IMU) 是测量姿势摆动的有效仪器,但它们检测健康老年人因视觉剥夺而产生的变化的能力需要进一步研究。我们检查了 IMU 传感器衍生的姿势摆动测量值与健康老年人在不同眼部条件下的力板相比的有效性和关系(32 名女性,33 名男性)。我们比较了质心和压力中心 (CoM 和 CoP) 衍生的总长度、均方根 (RMS) 距离、平均速度和 95%置信区间椭圆面积 (95%CI 椭圆面积) 的关系。此外,我们检查了 IMU 传感器在区分睁眼 (EO) 和闭眼 (EC) 条件方面与力板的关系。几乎所有变量都发现仪器和眼部条件存在显著影响。总体而言,EO 和 EC 变量在仪器内(力板 r 值为 0.38 至 0.78;IMU 传感器 r 值为 0.36 至 0.69)以及仪器间(r 值为 0.50 至 0.88)的相关性为中度至高度相关。RMS 距离和 95%CI 椭圆面积的 EC:EO 比值在仪器之间没有差异,而总长度(=0.973)和平均速度(=0.703)之间存在显著差异。仪器之间的比值相关系数范围从中度(r=0.65)到高度(r=0.87)。与力板相比,IMU 传感器为客观姿势摆动评估提供了一种经济实惠、有效的替代方案。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/3b20/11591754/030fc8e7a8a7/biosensors-14-00529-g001a.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验