Suppr超能文献

自动超声瞳孔评估:一种新的神经眼科评估工具。

Automated ultrasonographic pupillary assessment: a new neuro-ophthalmological assessment tool.

作者信息

Chen Weiting, Tang Min, Jin Yaling, Chen Lihui, Lin Jiuzhou, Yang Xuelin

机构信息

Department of Emergency Medicine, The First People's Hospital of Linhai, Taizhou, China.

Department of Emergency Medicine, Lishui Central Hospital, Lishui, China.

出版信息

J Int Med Res. 2024 Dec;52(12):3000605241299927. doi: 10.1177/03000605241299927.

Abstract

OBJECTIVE

To evaluate correlations between ultrasonographic pupillary assessment (UPA) and automated UPA in neurocritical patients.

METHODS

This was a prospective, observational study of 20 adult patients admitted to the intensive care unit with neurological pathology. Between UPA and automated UPA, 40 pupillary measurements were made. The time required to conduct UPA and automated UPA on a single patient was consistently < 3 minutes. Automated UPA required a markedly shorter operational time than UPA.

RESULTS

There were strong positive associations between UPA and automated UPA measurements, with the following correlations: right eye at rest,  = 0.9973; left eye at rest,  = 0.9989; right eye post-pupillary light response,  = 0.9975; left eye post-pupillary light response,  = 0.9955. Bland-Altman analyses confirmed the strong agreement between the two methods under both conditions, with most measurements falling within the limits of agreement. Both the right and left pupillary diameters at rest demonstrated consistency, and the post-pupillary light response measurements for both eyes indicated good agreement, with a few outliers.

CONCLUSIONS

Automated UPA was strongly correlated with UPA in neurocritical patients. Automated UPA is a rapid, feasible, and noninvasive method that enables the precise evaluation of pupillary size and light response.

摘要

目的

评估神经重症患者的超声瞳孔评估(UPA)与自动UPA之间的相关性。

方法

这是一项对20例入住重症监护病房的患有神经病理学疾病的成年患者进行的前瞻性观察研究。在UPA和自动UPA之间进行了40次瞳孔测量。对单个患者进行UPA和自动UPA所需的时间始终<3分钟。自动UPA所需的操作时间明显短于UPA。

结果

UPA与自动UPA测量之间存在强正相关,相关性如下:右眼静息时,=0.9973;左眼静息时,=0.9989;右眼瞳孔对光反射后,=0.9975;左眼瞳孔对光反射后,=0.9955。Bland-Altman分析证实了两种方法在两种情况下都有很强的一致性,大多数测量值落在一致性界限内。静息时左右瞳孔直径均显示出一致性,双眼瞳孔对光反射测量结果显示出良好的一致性,仅有少数离群值。

结论

在神经重症患者中,自动UPA与UPA密切相关。自动UPA是一种快速、可行且无创的方法,能够精确评估瞳孔大小和光反射。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/77f1/11618917/e30c825e2dbd/10.1177_03000605241299927-fig1.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验