Neeley W E, O'Classen K, Gruber M
Clin Chem. 1985 Jan;31(1):139-42.
Inexpensive improvements in continuous-flow analytical apparatus can eliminate some perplexing inconsistencies and limitations in the use of the Technicon SMAC instrument for measuring the activity in serum of aspartate aminotransferase (EC 2.6.1.1) in the presence of pyridoxal 5'-phosphate. In addition, a calibration procedure based on values for this enzyme obtained by a modified IFCC method can be used to calibrate both the Technicon SMAC and the Du Pont aca instruments to produce excellent correlation between the two.