Sinak L J, Liu Y H, Block M, Mair D D, Julsrud P R, Hoffman E A, Hagler D J, Seward J B, Ritman E L
J Am Coll Cardiol. 1985 Jan;5(1 Suppl):70S-76S. doi: 10.1016/s0735-1097(85)80146-7.
The Dynamic Spatial Reconstructor is a unique high speed volume imaging X-ray scanner based on computed tomographic principles. It has several potential advantages over conventional angiographic methods, including reduced invasion, reduced rate of false negative results and increased accuracy of measurements of structure and function. To evaluate the utility of the Dynamic Spatial Reconstructor in the investigation of congenital heart disease, scanning was performed in several pediatric patients with pulmonary valve atresia. Early results show that three-dimensional images of the cardiac chambers and intrathoracic vessels can be acquired and displayed with the system. All the information necessary for quantitative analysis of the pulmonary arteries can be obtained with a single scan involving injection of 0.6 to 2.0 cc/kg of contrast agent and radiation exposure of 0.54 to 0.95 rad/s over 5 to 8 seconds.
动态空间重建仪是一种基于计算机断层扫描原理的独特的高速容积成像X射线扫描仪。与传统血管造影方法相比,它具有几个潜在优势,包括侵入性降低、假阴性结果率降低以及结构和功能测量的准确性提高。为了评估动态空间重建仪在先天性心脏病研究中的效用,对几名患有肺动脉瓣闭锁的儿科患者进行了扫描。早期结果表明,该系统可以采集并显示心腔和胸内血管的三维图像。通过单次扫描,注入0.6至2.0 cc/kg的造影剂,在5至8秒内辐射暴露为0.54至0.95 rad/s,就可以获得肺动脉定量分析所需的所有信息。