Zhang Xin, Xie Jixiong, Tan Yuhang, Su Ting, Zhu Jiongtao, Cui Han, Xia Dongmei, Zheng Hairong, Liang Dong, Ge Yongshuai
Research Center for Advanced Detection Materials and Medical Imaging Devices, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong, China.
University of Chinese Academy of Sciences, Beijing, China.
Med Phys. 2025 Mar;52(3):1500-1514. doi: 10.1002/mp.17567. Epub 2024 Dec 17.
Recently, the popularity of dual-layer flat-panel detector (DL-FPD) based dual-energy cone-beam CT (CBCT) imaging has been increasing. However, the image quality of dual-energy CBCT remains constrained by the Compton scattered x-ray photons.
The objective of this study is to develop a novel scatter correction method, named e-Grid, for DL-FPD based CBCT imaging.
In DL-FPD, a certain portion of the x-ray photons (mainly low-energy [LE] primary and scattered photons) passing through the object are captured by the top detector layer, while the remaining x-ray photons (mainly high-energy [HE] primary and scattered photons) are collected by the bottom detector layer. A linear signal model was approximated between the HE primary and scatter signals and the LE primary and scatter signals. Physical calibration experiments were performed on cone beam and fan beam to validate the aforementioned signal model via linear fittings. Monte Carlo (MC) simulations of a 10 cm diameter water phantom were conducted on GATE at first to verify this newly developed scatter estimation method. In addition, physical validation experiments of water phantom, head phantom, and abdominal phantom were carried out on a DL-FPD based benchtop CBCT imaging system. The image non-uniformity (NU), which represents the relative difference between the center and the edges of CT images, was measured to quantify the reduction of image shading artifacts. Finally, multi-material decomposition was conducted.
The MC results, CBCT images and line profiles, showed that the newly proposed e-Grid approach was able to accurately predict the scatter distributions in both shape and intensity. As a result, uniform CBCT images that are close to the scatter artifact-free reference images can be obtained. Moreover, the physical experiments demonstrated that the e-Grid method can greatly reduce the shading artifacts in both LE and HE CBCT images acquired from DL-FPD. Results also demonstrated that the e-Grid method is effective for varied objects that having different diameters (from 10 to 28 cm). Quantitatively, the NU value was reduced by over 77% in the LE CBCT image and by over 66% in the HE CBCT image on average. As a consequence, the accuracy of the decomposed multi-material bases, iodine and gadolinium, was substantially improved.
The Compton scattered x-ray signals could be significantly reduced using the proposed e-Grid method for DL-FPD based dual-energy CBCT imaging systems.
近年来,基于双层平板探测器(DL-FPD)的双能锥束CT(CBCT)成像越来越受欢迎。然而,双能CBCT的图像质量仍受康普顿散射X射线光子的限制。
本研究的目的是为基于DL-FPD的CBCT成像开发一种名为e-Grid的新型散射校正方法。
在DL-FPD中,穿过物体的一部分X射线光子(主要是低能[LE]原发射线光子和散射光子)被顶部探测器层捕获,而其余的X射线光子(主要是高能[HE]原发射线光子和散射光子)则由底部探测器层收集。在HE原发射线信号与散射信号以及LE原发射线信号与散射信号之间近似建立线性信号模型。在锥束和扇束上进行物理校准实验,通过线性拟合验证上述信号模型。首先在GATE上对直径10 cm的水模进行蒙特卡罗(MC)模拟,以验证这种新开发的散射估计方法。此外,在基于DL-FPD的台式CBCT成像系统上对水模、头部模型和腹部模型进行了物理验证实验。测量图像不均匀性(NU),其代表CT图像中心与边缘之间的相对差异,以量化图像阴影伪影的减少。最后,进行多材料分解。
MC结果、CBCT图像和线轮廓表明,新提出的e-Grid方法能够准确预测散射分布的形状和强度。因此,可以获得接近无散射伪影参考图像的均匀CBCT图像。此外,物理实验表明,e-Grid方法可以大大减少从DL-FPD获取的LE和HE CBCT图像中的阴影伪影。结果还表明,e-Grid方法对不同直径(从10到28 cm)的各种物体均有效。定量地说,LE CBCT图像中的NU值平均降低了77%以上,HE CBCT图像中的NU值平均降低了66%以上。因此,分解后的多材料基碘和钆的准确性得到了显著提高。
使用所提出的e-Grid方法,基于DL-FPD的双能CBCT成像系统中的康普顿散射X射线信号可以显著减少。