Maguen E, Nesburn A B, Jackman N, Macy J I
Am J Ophthalmol. 1985 Feb 15;99(2):170-2. doi: 10.1016/0002-9394(85)90226-0.
The haptic of a semiflexible anterior chamber intraocular lens (Kelman Quadraflex) broke inside the eye of a 70-year-old man. No history of major eye or head trauma could be elicited from the patient. Clinical manifestations included inferior corneal edema progressing toward the optical axis resulting from a free-floating haptic segment of the lens implant inside the anterior chamber. A significant decrease in endothelial density was documented over a short period of time. The surgical management of this case included removal of the free-floating haptic segment, removal of the lens implant, penetrating keratoplasty, and vitrectomy. Final visual acuity was 20/70.