Suppr超能文献

半硬性人工晶状体植入物破裂。

Broken semiflexible intraocular lens implant.

作者信息

Maguen E, Nesburn A B, Jackman N, Macy J I

出版信息

Am J Ophthalmol. 1985 Feb 15;99(2):170-2. doi: 10.1016/0002-9394(85)90226-0.

Abstract

The haptic of a semiflexible anterior chamber intraocular lens (Kelman Quadraflex) broke inside the eye of a 70-year-old man. No history of major eye or head trauma could be elicited from the patient. Clinical manifestations included inferior corneal edema progressing toward the optical axis resulting from a free-floating haptic segment of the lens implant inside the anterior chamber. A significant decrease in endothelial density was documented over a short period of time. The surgical management of this case included removal of the free-floating haptic segment, removal of the lens implant, penetrating keratoplasty, and vitrectomy. Final visual acuity was 20/70.

摘要

一枚半柔性前房型人工晶状体(Kelman Quadraflex)的触觉部在一名70岁男性患者眼内断裂。患者无重大眼部或头部外伤史。临床表现包括因前房内晶状体植入物的触觉部游离片段导致角膜水肿向下朝着视轴进展。短期内记录到内皮细胞密度显著下降。该病例的手术处理包括取出游离的触觉部片段、取出晶状体植入物、穿透性角膜移植术和玻璃体切除术。最终视力为20/70。