The influence of inner wall filamentousness on the performance of small- and large-caliber arterial grafts.

作者信息

Wu H D, Zammit M, Sauvage L R, Streicher M D

出版信息

J Vasc Surg. 1985 Mar;2(2):255-62.

PMID:3974011
Abstract

Dacron prostheses with differing degrees of inner wall filamentousness were implanted in the canine descending thoracic aorta for 56 days (n = 14) and the carotid arteries for 21 days (n = 40). In the thoracic aorta the healed inner capsule of 8 mm internal velour grafts was approximately three times as thick as the inner capsule of external velour knitted prostheses. However, the early patency and rapidity of healing of these two types of prostheses were essentially the same. In the carotid arteries the patency of the 4 mm external velour prostheses was greater than that of the internal velour prostheses and the mean thrombus-free surface area of the external velour grafts was significantly greater than that of the internal velour grafts. These observations suggest that an internal velour surface is not deleterious to early patency in a large-caliber, high-flow artery but that it may be in a small-caliber, low-flow artery.