肺密度对¹²⁵I剂量分布的影响。

Lung density effect on 125I dose distribution.

作者信息

Prasad S C, Bassano D A, Perg J G

出版信息

Med Phys. 1985 Jan-Feb;12(1):99-100. doi: 10.1118/1.595742.

Abstract

Perturbations in 125I implant dose distribution due to lung tissue variations have been investigated. Dose correction factors for a point source, a planar, and a volume implant have been calculated using a model which accounts for the changes in primary photon attenuation and in buildup factor when the medium is lung rather than water. Results of our calculation show that the change in dose is about 9% and 20% in the core and the periphery, respectively, of a representative implanted volume whose density is 0.25 relative to water.

摘要

已对因肺组织差异导致的¹²⁵I植入剂量分布扰动进行了研究。使用一个模型计算了点源、平面和体积植入的剂量校正因子,该模型考虑了介质为肺而非水时初级光子衰减和积累因子的变化。我们的计算结果表明,对于密度相对于水为0.25的代表性植入体积,其核心和周边的剂量变化分别约为9%和20%。