Lewandowska Agata, Borowicz Dorota, Bielęda Grzegorz, Nowotka Oliwia, Kazior Maksymilian, Krupka Justyna, Kanikowski Marek, Kozak Maciej
Medical Physics Department, Affidea Cancer Treatment Center, Poznań, Poland.
Department of Biomedical Physics, Faculty of Physics, Adam Mickiewicz University, Poznań, Poland.
Rep Pract Oncol Radiother. 2024 Dec 4;29(5):588-599. doi: 10.5603/rpor.102819. eCollection 2024.
The purpose of this study is to determine the effect of the type of I-125 radioactive source on dose distribution in the planning process of ultra-low dose rate (uLDR) prostate brachytherapy.
7 patients who had undergone brachytherapy in our center were included in the study. Dose in five geometrical points were analyzed for 12 types of implants that are available on the market. The plans were originally calculated for S17plus implant model (Eckert & Ziegler Medical). Point dose calculations were performed using RadCalc (LAP) software.
The differences in doses for individual points were significant. The largest differences were observed at the point located in the center of the patients' urethra and between BestMedical 2301 (highest doses for each point) and IsoStar (lowest doses) implants.
The choice of implant manufacturer strongly influences what dose distribution will be obtained for the prostate. It is possible to obtain satisfactory plans that meet the dosimetric criteria for each type of implant, but the positioning of each source will vary significantly.
本研究的目的是确定在超低剂量率(uLDR)前列腺近距离放射治疗的计划过程中,I-125放射源类型对剂量分布的影响。
本研究纳入了7例在本中心接受过近距离放射治疗的患者。对市场上现有的12种植入物模型,分析了五个几何点处的剂量。这些计划最初是针对S17plus植入物模型(埃克特&齐格勒医疗公司)计算的。使用RadCalc(LAP)软件进行点剂量计算。
各点剂量差异显著。在患者尿道中心的点以及BestMedical 2301(各点剂量最高)和IsoStar(剂量最低)植入物之间观察到最大差异。
植入物制造商的选择对前列腺所获得的剂量分布有很大影响。对于每种植入物类型,都有可能获得符合剂量学标准的满意计划,但每个放射源的定位会有显著差异。