Flow cytometric enumeration of absolute lymphocyte number in peripheral blood using two parameters of light scatter.

作者信息

Smart Y C, Cox J, Murphy B, Enno A, Burton R C

出版信息

Cytometry. 1985 Mar;6(2):172-4. doi: 10.1002/cyto.990060214.

Abstract

A method was developed to measure the absolute lymphocyte count (ALC) of whole blood using the Spectrum III automated flow cytometer. Ninety-nine samples of human peripheral blood were analysed on the Spectrum and the Coulter Counter S Plus II, to allow for comparison of the two machines. Regression analysis was used to test the extent of agreement between the sets of measurements on the two machines. The results demonstrated that the slope of the regression line was not significantly different from one, indicating a high level of correlation between Spectrum and Coulter ALC's. However, the mean difference between Coulter and Spectrum ALC's was not equal to zero, with the Spectrum giving counts approximately 10% lower than those of the Coulter machine. This is attributed to the different ways by which the two machines define a lymphocyte, the Spectrum III by two parameters of light scatter and the Coulter S Plus II by the single parameter of cell volume.

摘要

开发了一种使用Spectrum III自动流式细胞仪测量全血绝对淋巴细胞计数(ALC)的方法。对99份人类外周血样本在Spectrum和库尔特计数器S Plus II上进行分析,以便对两台机器进行比较。使用回归分析来检验两台机器测量结果之间的一致程度。结果表明,回归线的斜率与1没有显著差异,表明Spectrum和库尔特ALC之间具有高度相关性。然而,库尔特和Spectrum ALC之间的平均差异不等于零,Spectrum给出的计数比库尔特机器的计数低约10%。这归因于两台机器定义淋巴细胞的不同方式,Spectrum III通过两个光散射参数,而库尔特S Plus II通过细胞体积的单一参数。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验