球囊扩张式经导管主动脉瓣置换术后人工心脏瓣膜扩张的新型实时荧光透视评估方法

Novel Real-Time Fluoroscopic Assessment Method of THV Expansion Following Balloon-Expandable TAVR.

作者信息

Husain Ali, Tzimas Georgios, Akodad Mariama, Jelisejevas Julius, Offen Sophie, Millar Kevin, Leipsic Jonathon A, Blanke Philipp, Wood David A, Sellers Stephanie L, Webb John G, Meier David, Sathananthan Janarthanan

机构信息

Centre for Cardiovascular Innovation (A.H., J.J., S.O., K.M., J.A.L., P.B., D.A.W., S.L.S., J.G.W., J.S.), University of British Columbia, Vancouver, Canada.

Centre for Heart Valve Innovation, St. Paul's Hospital (A.H., J.J., S.O., K.M., J.A.L., P.B., D.A.W., S.L.S., J.G.W., J.S.), University of British Columbia, Vancouver, Canada.

出版信息

Circ Cardiovasc Interv. 2025 Feb;18(2):e014617. doi: 10.1161/CIRCINTERVENTIONS.124.014617. Epub 2025 Jan 24.

Abstract

BACKGROUND

Transcatheter heart valve (THV) underexpansion after transcatheter aortic valve replacement may be associated with worse outcomes. THV expansion can be assessed fluoroscopically using a pigtail for calibration; however, the accuracy of this technique specific to transcatheter aortic valve replacement is unknown. We assessed the accuracy and reproducibility of a novel fluoroscopic method to assess THV expansion using the THV commissural post for calibration.

METHODS

Patients who underwent transcatheter aortic valve replacement with a SAPIEN 3 (S3) THV had documented 3 cusp and cusp overlap views, and post-implant computed tomography was identified. THV expansion was fluoroscopically assessed in the 3 cusp and cusp overlap views using the S3 commissural post height and the pigtail as a reference for calibration. The correlation between the 2 methods and computed tomography was evaluated.

RESULTS

Forty patients were included. On the bench, the commissural post height measured 3.3, 3.5, 4.0, and 4.5 mm for the 20 mm S3, 23 mm S3, 26 mm S3, and 29 mm S3, respectively. The Pearson correlation coefficient (r) with computed tomography for measuring the inflow, mid-portion, and outflow THV diameter was 0.98, 0.97, and 0.98 for the commissural post height method and 0.82, 0.81, and 0.78 for the pigtail method, respectively. Unlike the pigtail method, the correlation between the commissural post height method and computed tomography remained strong across all THV sizes and in both the 3 cusp and cusp overlap views.

CONCLUSIONS

The commissural post height method is a novel real-time fluoroscopic tool that has the potential to assess THV expansion and guide further optimization after balloon-expandable transcatheter aortic valve replacement.

摘要

背景

经导管主动脉瓣置换术后经导管心脏瓣膜(THV)扩张不足可能与更差的预后相关。可通过使用猪尾导管进行校准,在荧光透视下评估THV扩张;然而,这种特定于经导管主动脉瓣置换术的技术的准确性尚不清楚。我们评估了一种使用THV连合柱进行校准来评估THV扩张的新型荧光透视方法的准确性和可重复性。

方法

接受SAPIEN 3(S3)THV经导管主动脉瓣置换术的患者有记录的3个瓣叶和瓣叶重叠视图,并确定了植入后计算机断层扫描。在3个瓣叶和瓣叶重叠视图中,使用S3连合柱高度和猪尾导管作为校准参考,通过荧光透视评估THV扩张。评估了这两种方法与计算机断层扫描之间的相关性。

结果

纳入40例患者。在实验台上,20mm S3、23mm S3、26mm S3和29mm S3的连合柱高度分别测量为3.3mm、3.5mm、4.0mm和4.5mm。连合柱高度法测量流入、中部和流出THV直径与计算机断层扫描的Pearson相关系数(r)分别为0.98、0.97和0.98,猪尾导管法分别为0.82、0.81和0.78。与猪尾导管法不同,连合柱高度法与计算机断层扫描之间的相关性在所有THV尺寸以及3个瓣叶和瓣叶重叠视图中均保持很强。

结论

连合柱高度法是一种新型实时荧光透视工具,有可能在球囊扩张式经导管主动脉瓣置换术后评估THV扩张并指导进一步优化。