Hayase Justin H, Jain Rohit, Lux Robert, Mori Shumpei, Ajijola Olujimi A, Shivkumar Kalyanam, Bradfield Jason S
UCLA Cardiac Arrhythmia Center, Ronald Reagan UCLA Medical Center, Los Angeles, California.
UCLA Cardiac Arrhythmia Center, Ronald Reagan UCLA Medical Center, Los Angeles, California.
Heart Rhythm. 2025 Aug;22(8):2099-2110. doi: 10.1016/j.hrthm.2025.01.033. Epub 2025 Jan 31.
In ventricular arrhythmia ablation procedures, traditional voltage mapping calculates overall peak-to-peak measurements. However, this methodology incorporates multiple signal components that do not distinguish near-field vs far-field components.
This study aimed to determine how the use of local bipolar measurements as identified by the first derivative of voltage over time (dV/dT) affects traditional voltage mapping characteristics.
Percutaneous endocardial and epicardial electroanatomic mapping was performed in a porcine myocardial infarction model (n = 5) with a multipolar and a point-by-point catheter. Electrograms were examined offline based on standard bipolar and unipolar voltage values and then reanalyzed by a novel algorithm that measures only the voltage magnitude pertaining to the maximum absolute dV/dT within a given electrogram. Computed tomography, gross pathology, and histopathology features were examined.
There were 25,114 bipolar mapping points across multipolar catheter maps and 6317 mapping points across point-by-point catheter maps. A difference in calculated voltage occurred in >80% of all mapping points, with all changes using dV/dT methodology resulting in lower bipolar voltage values compared with traditional methods. There was a categorical change in standard voltage (scar, border zone, normal) in >7% of all points. Unipolar voltage calculated by local dV/dT also resulted in a lower value in >90% of all mapping points. Histologic examination of discrepant regions revealed patterns of diffuse fibrosis.
In a porcine infarct model, the use of dV/dT to identify bipolar voltage results in lower measured values in >80% of all mapping points compared with traditional peak-to-peak measurement. This methodology may more accurately identify local tissue properties by selection of near-field components within substrates implicated in ventricular arrhythmias.
在室性心律失常消融手术中,传统的电压标测计算的是整体峰峰值测量值。然而,这种方法包含多个信号成分,无法区分近场和远场成分。
本研究旨在确定使用电压随时间的一阶导数(dV/dT)识别的局部双极测量值如何影响传统电压标测特征。
在猪心肌梗死模型(n = 5)中,使用多极导管和逐点导管进行经皮心内膜和心外膜电解剖标测。基于标准双极和单极电压值对心电图进行离线检查,然后通过一种仅测量给定心电图内与最大绝对dV/dT相关的电压幅值的新算法进行重新分析。检查计算机断层扫描、大体病理学和组织病理学特征。
多极导管标测中有25114个双极标测点,逐点导管标测中有6317个标测点。在所有标测点中,>80%的计算电压存在差异,与传统方法相比,使用dV/dT方法的所有变化均导致双极电压值降低。在所有点中,>7%的标准电压(瘢痕、边界区、正常)发生了类别变化。通过局部dV/dT计算的单极电压在所有标测点中>90%也导致较低的值。对差异区域的组织学检查显示出弥漫性纤维化模式。
在猪梗死模型中,与传统的峰峰值测量相比,使用dV/dT识别双极电压导致在所有标测点中>80%的测量值较低。这种方法可能通过选择与室性心律失常相关的基质内的近场成分,更准确地识别局部组织特性。